Printing
N-560 Operator’s Manual
59
The last data field in the top line indicates the upper and lower alarm
limits for %SpO
2
and for the pulse rate (PR), and the
SatSeconds
alarm
setting (OFF). The
SatSeconds
setting may be OFF, 10, 25, 50, or 100
depending on the
SatSeconds
alarm setting. In the example above, the low
alarm limit for SpO
2
is 85% and the upper alarm limit is 100%. Pulse Rate
alarm limits are 40 to 170 bpm.
SatSeconds
alarm limit is set to off.
N-560 Mode
The N-560 mode is ADULT.
Data Column Headings
Actual column headings are in the third row of the column heading line.
Patient data that is presented in the chart, from left to right, are the time
that the patient data on the row was obtained, the current %SpO
2
value
being measured, the current Pulse Rate in beats per minute (bpm), the
current Pulse Amplitude (PA), and the operating status of the N
-
560.
Time
The Time column represents the N
-
560 real-time clock. Refer to the
N
-
560 service manual to set the N
-
560 real-time clock.
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Page 12: ...Introduction 4...
Page 22: ...Description of Controls Indicators and Symbols 14...
Page 50: ...Using the N 560 42...
Page 54: ...N 560 Trend 46...
Page 70: ...Printing 62...
Page 76: ...Sensors and Accessories 68...
Page 82: ...Operator s Menu 74...
Page 90: ...Troubleshooting 82...
Page 98: ...Technical Information 90...
Page 118: ...Specifications 110...
Page 122: ...Index 114...
Page 123: ......