Using the N-560
22
Procedure
1. Turn on the N
-
560 by pressing and holding the
Power On/Off
button
for more than one second.
2. The N-560 displays/sounds:
3. The N-560 automatically starts the Power
-
On Self
-
Test (POST),
which tests N-560 circuitry and functions.
CAUTION: During POST (immediately after power-up), confirm
that all display segments and indicators light, and the speaker sounds
a 1-second pass tone.
4. While performing POST, the self
-
test display appears for
approximately 2 to 4 seconds. During this time:
Display
Sound
5 (in pulse rate left window)
one beep tone
6 (in pulse rate center window)
one beep tone
0 (in pulse rate right window)
one beep tone
n (in SpO
2
left window)
none
n (in SpO
2
center window)
none
Summary of Contents for OXIMAX N-560