Description of Controls, Indicators, and Symbols
N-560 Operator’s Manual
13
Description of Audible Indicators
Following are descriptions of N
-
560 audible indicators.
Power
-
On Self
-
Test Pass A 1
-
second tone indicating that the N
-
560 has been
turned on and has successfully completed the
power
-
on self
-
test
Valid Button Press
A short, medium
-
pitched tone indicating that an
appropriate button has been pressed
Invalid Button Press
A short, low
-
pitched tone indicating that a button
has been pressed that is not appropriate for the
current state of the N-560
High Priority Alarm
A high
-
pitched, fast
-
pulsating tone indicating loss
of pulse
Medium Priority Alarm
A medium
-
pitched, normal
-
pulsating tone
indicating an SpO
2
or pulse rate limit violation
Low Priority Alarm
A low
-
pitched, slow
-
pulsating tone indicating a
sensor disconnect, low battery, or N-560 failure
Alarm Silence Reminder Three beeps that sound approximately every
3 minutes when alarms are silenced with the alarm
silence duration set to OFF
Pulse Beep
A single beep sounds for each detected pulse. The
pitch changes as monitored SpO
2
values increase or
decrease.
Volume Setting Tone
A continuous tone that is used to adjust the alarm
volume
Confirmation Tone
Three beeps sound to indicate that default settings
have been saved or reset to factory defaults or trend
data has been deleted
Summary of Contents for OXIMAX N-560
Page 1: ...C M Y CM MY CY CMY K Ops ai 1 24 06 9 31 41 AM Ops ai 1 24 06 9 31 41 AM...
Page 12: ...Introduction 4...
Page 22: ...Description of Controls Indicators and Symbols 14...
Page 50: ...Using the N 560 42...
Page 54: ...N 560 Trend 46...
Page 70: ...Printing 62...
Page 76: ...Sensors and Accessories 68...
Page 82: ...Operator s Menu 74...
Page 90: ...Troubleshooting 82...
Page 98: ...Technical Information 90...
Page 118: ...Specifications 110...
Page 122: ...Index 114...
Page 123: ......