Specifications
96
Table 6: Tone Definition
Audible Indicator
Parameter
Value
Alarm Silence
Reminder
Volume level
Not changeable
Pitch (±30 Hz)
784 Hz
Pulse width (±20 ms)
150 ms
Pulse repetition interval
(±20 ms) (double burst)
150 ms
Repetitions
3
Confirmation of Button
Pressed
Volume level
Not changeable
Pitch (±30 Hz)
784 Hz
Pulse width (±20 ms)
150 ms
Pulse repetition interval
(±20 ms) (double burst)
150 ms
Repetitions
3
High Priority Alarm
Volume level
Adjustable alarm
volume
Pitch (±30 Hz)
932 Hz
Pulse width (±20 ms)
Nellcor = 255 ms
IAW IEC 60601-1-8
= 120 ms
IAW EN 475 =
150 ms
Pulse repetition interval
(±20 ms) (double burst)
Nellcor = 320 ms
IAW IEC 60601-1-8
= 6940 ms
IAW EN 475 =
7500 ms
Repetitions
Continually