Using the Monitor
N-550
29
3. The Alarm Silence indicator will flash, indicating that the alarm
sounds are disabled. The monitor will sound three beeps
approximately every three minutes to warn the user that the alarm
sound has been silenced.
Verify Patient Settings
With the monitor in the normal monitoring mode:
1. Press the SpO
2
Alarm Limit button to view the current %SpO
2
upper alarm limit.
2. Press the SpO
2
Alarm Limit button twice to view the current
%SpO
2
lower alarm limit.