N-550
83
S p e c i f i c a t i o n s
Performance
Measurement Range
SpO
2
1% to 100%
Pulse Rate
0 and 20 beats per minute (BPM) to
250 BPM
Perfusion Range
0.03% to 20%
Accuracy and Motion Tolerance
Saturation
Without Motion - Adults
1
70 to 100% ±2 digits
Without Motion - Neonate
1
70 to 100% ±3 digits
With Motion - Adults and
Neonates
2
70 to 100% ±3 digits
Low Perfusion
3
70 to 100% ±2 digits
Pulse Rate
Without Motion
3
20 to 250 bpm ±3 digits
With Motion
normal physiological range
(55 - 125 bpm) ±5 digits
Low Perfusion
3
20 to 250 bpm ±3 digits