N-180 Pulse Oximeter Service Manual
8-2
ENVIRONMENTAL REQUIREMENTS
Temperature
Monitor
50-104˚ F (10-40˚C) operating
32-122˚ F (0-50˚C) storage
Sensor
Within physiologic range 82.4–107.6˚F (28–42˚C) for accurate measurement
Humidity
Any humidity/temperature combination without condensation
Altitude
0–10,000 ft (3,048 m)
ELECTRICAL CHARACTERISTICS
Protective Class
Class I
Mains-supplied unit using a protective ground
Degree of Protection
Type BF
Patient electrically isolated
Voltage
100–120 VAC
±
10%, 50–60 Hz
or 220–240 VAC
±
10%, 50-60 Hz
Power Consumption
Maximum rating
:
25 VA
Battery
Type
Lead -acid battery pack, 1.9 AH
Battery Life
Typically
6 hours for a new, fully charged battery. Capacity decreases with
age.
Recharge Period
14 hours minimum