English-24

STAND-ALONE CALIBRATION

Stand alone calibration can recalibrate the display without using a computer by the front sensor or connecting an external 
sensor. There are 5 functions in STAND-ALONE CALIBRATION.

• 

Self Calibration:

 Calibrates the front sensor based on the USB optical sensor. Alternatively, manually adjust the integrated 

front sensor with an external near range luminance sensor reference, which is traceable to a primary standards laboratory.

• 

Copy Calibration:

 Copies the luminance from one display to one or more additional displays.

• 

Gamma Adjust: 

Adjusts Gamma value.

• 

DICOM Measurement:

 Measures DICOM response.

• 

AMBIENT SENSOR CALIBRATION: 

To manually adjust the integrated front sensor with an external illuminance sensor 

reference, which is traceable to a primary standards laboratory. 

NOTE:

 

Stand before proper Stand-Alone calibration can be performed, the display should warm-up for a minimum of 
30 minutes. If Stand-Alone calibration is started before the monitor is warmed up, a warning will appear on-screen 

(Figure A.1)

.

When Stand-Alone Calibration is executed, the message window will appear (

Figure A.3

) and the screen becomes dark for 

about 10 seconds. Then ZERO CALIBRATION will start automatically.

NOTE:

 

Please only use an USB optical sensor according to our specifi cation.
Refer to the KEY MAP 

(Figure A.2)

 to adjust the settings for calibration.

Stand-Alone calibration should be performed in either portrait or landscape orientation.

NOTE:

 

To start calibration without a computer, perform the following steps.
When the “NO SIGNAL” message appears on screen, press the “LEFT” and then “RIGHT” buttons and hold them 
down simultaneously. The CALIBRATION menu will appear.

NOTE:

 

When Stand-Alone Calibration is executed, the message window will appear (

Figure A.3

) and the screen becomes 

dark about 10 seconds. It is not a malfunction.

Figure A.2

KEY MAP

UP/DOWN:

 

Changes from setting to setting

LEFT/RIGHT:

  Changes the setting selection 

 

(ie...SELF or COPY)

SELECT:

 

Moves to the next step of calibration

EXIT:

 

Moves back one step of calibration

Figure A.1

Figure A.3

CAUTION:  Please use a properly-managed USB optical sensor. If there is any doubt about the result of the Stand- 

 

  Alone calibration result, please contact your supplier. 

Summary of Contents for MD211G3-R

Page 1: ...Medical Display MD211G3 INSTALLATION MAINTENANCE GUIDE...

Page 2: ...nit This symbol alerts the user that important literature concerning the operation and maintenance of this unit has been included Therefore it should be read carefully in order to avoid any problems P...

Page 3: ...REFER SERVICING TO QUALIFIED SERVICE PERSONNEL CAUTION Power Cord Important Information CAUTION Please use the power cord provided with this monitor in accordance with the table below If a power cord...

Page 4: ...ms is not recommended if cleaning of the air intakes from dust and textile lint in regular time intervals cannot be performed by the user maximum interval six months This device is equipped with an in...

Page 5: ...at interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and...

Page 6: ...ot install where the monitor will be exposed to continual vibration If monitor or glass is broken do not come in contact with the liquid crystal and handle with care To prevent damage to the LCD monit...

Page 7: ...optimum performance allow 30 minutes for the monitor to warm up Adjust the monitor height so that the top of the screen is at or slightly below eye level Your eyes should look slightly downward when...

Page 8: ...DisplayPort cable CD ROM x 2 Power Cord 2 Quick Reference Guide USB Cable Remember to save your original box and packing material to transport or ship the monitor 1 Dual link DVI cable 2 Type and num...

Page 9: ...nnect the DisplayPort cable to the connector of the display controller in your system Figure A 2 Figure A 1 Figure A 2 DisplayPort cable NOTE 1 Please use a DisplayPort cable with a DisplayPort logo 2...

Page 10: ...sub monitors can be daisy chained Figure C 1 Highest Stand Position Power cord 30 Tilt Figure C 1a A Type B Type A Type B Type Sensor Port 1 DVI D Cable DisplayPort Cable USB Upstream Port USB Downst...

Page 11: ...switch on the top Portrait or left side Landscape of the monitor Figure E 1 Turn on the computer NOTE If you have any problems please refer to the Troubleshooting section of this Manual Figure E 1 Po...

Page 12: ...screen place hands on each side of the monitor and turn clockwise from Landscape to Portrait or counter clockwise from Portrait to Landscape Figure R 1 To rotate the OSD menu between landscape and por...

Page 13: ...down on a desk or table with the front sensor situated off the edge Figure S 2 5 Slide the cable cover up NOTE The cable cover is not removable 6 Place hand on the back side of the monitor and lift up...

Page 14: ...xits any OSD sub menu Exits the OSD Control menu Shows OSD menu Shows PICTURE MODE Selects the item Changes the input source Resets the OSD back to factory settings Shows the quick screen QA test patt...

Page 15: ...URE MODE Change to a picture mode which is most suitable for the type of content that is shown You can choose one of two possible PICTURE MODES See Advanced OSD menu Tag1 PICTURE MODE page 17 and Usin...

Page 16: ...tor will automatically return to normal mode QA TEST Provides a quick daily visual image quality check of this diagnostic monitor by just using a button on the monitor MENU Tools LANGUAGE The OSD cont...

Page 17: ...ves a warning when there is no Horizontal or Vertical Sync After power is turned on or when there is a change of input signal the No Signal message window will appear OUT OF RANGE This function gives...

Page 18: ...F ERROR result in is a quick indicator of the gray level quality against the ideal DICOM curve Depending on the QA standard which is used in your country different GSDF error limits may exist Please c...

Page 19: ...m by a monitor Gamma is defined by the relationship between the signal input voltage and the resulting intensity of the monitor s light output A perfect linear device would have a Gamma of 1 0 A Gamma...

Page 20: ...RIGHT to select FIRST When current video input signal is not present then the monitor searches for a video signal from a different video source If a video signal is supplied by the other video source...

Page 21: ...a screen will appear indicating the OSD controls are locked out There are three types of OSD LOCK OUT OSD LOCK OUT with no control To activate the OSD Lock Out function press SELECT then RIGHT button...

Page 22: ...y a sensor LANDSCAPE The preferred timing is fixed to landscape orientation 2048x1536 pixels PORTRAIT The preferred timing is fixed to portrait orientation 1536x2048 pixels Note If you have any proble...

Page 23: ...l display performance Active Display Area Landscape Horiz Vert Portrait Horiz Vert 433 2 mm 17 1 inches 324 9 mm 12 8 inches 324 9 mm 12 8 inches 433 2 mm 17 1 inches USB Hub I F Port Load Current USB...

Page 24: ...resolutions supported directly to your computer automatically optimizing display performance Intelligent Power Manager System Provides innovative power saving methods that allow the monitor to switch...

Page 25: ...mage NOTE As with all personal display devices NEC DISPLAY SOLUTIONS recommends using a moving screen saver at regular intervals whenever the screen is idle or turning off the monitor when not in use...

Page 26: ...pear on screen Figure A 1 When Stand Alone Calibration is executed the message window will appear Figure A 3 and the screen becomes dark for about 10 seconds Then ZERO CALIBRATION will start automatic...

Page 27: ...S 3 Press SELECT to go to the next section To select AUTO CALIB go to step 6 To select EXPERT CALIB go to step 7 AUTO CALIB Make sure that the USB optical sensor is connected to the sensor port 6 The...

Page 28: ...n 5 The procedure will ask for the USB optical sensor to be placed on the center of the display panel Figure C 4 Tilt the display panel approximately 5 backward and place the USB optical sensor in the...

Page 29: ...the advanced menu page 24 2 Select GAMMA ADJUST in the MODE selection Figure 1 3 Select the target UNIFORMITY Press SELECT to go to the next section 4 Set the target LUMINANCE cd m2 Figure 2 Press SEL...

Page 30: ...setting is 1 or 2 Please perform AMBIENT SENSOR CALIBRATION in the following cases The usage environment has changed AMBIENT SENSOR CALIBRATION requires an illuminance meter and a telescopic luminanc...

Page 31: ...aving measured press SELECT without changing the value 6 After the CALIBRATION SUCCEEDED message appears Figure AS 4 press SELECT You can set DATE and TIME by using control keys and press SELECT 7 To...

Page 32: ...t Signalling DPMS Standard signal is sent to the monitor the Energy Saving mode is activated The monitor enters a single Energy Saving mode Mode Power consumption LED colour Normal Operation Approx 10...

Reviews: