iv | Safety Information
Power Supply
For Customers in North America
EndoVue series monitors comply to the above standards only when used with the supplied hospital grade power
supply.
Note:
This power supply is a part of the medical device.
For Customers in China
This power supply is to be used with this monitor.
Power Cord
For Customers in North America
Use a hospital grade power cord with the correct plug for your power source.
The power cord is the only recognized disconnect device. Disconnect the power cord from the AC mains.
MEDICAL EQUIPMENT should be positioned so that the disconnect device is readily accessible.
Monitors should be powered from a center tapped circuit when used in the US at voltages over 120 volts.
For Customers in China
Use China Compulsory Certificate (CCC)-approved power cord.
For Customers in Japan
The supplied power cord is for use only with this monitor.
Grounding
This product is energized from an external electrical power source for Class 1 equipment. It is the responsibility of
the installer to test the product’s earth ground to verify that it complies with the hospital, local and national
impedance requirements.
A ground post is located on the back of the product to use for grounding the chassis of the unit. Any such ground
must be installed in accordance with applicable electrical codes. The ground post location is shown in the
EndoVue Connector Panel diagram on page 7.
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle
marked “Hospital Only” or “Hospital Grade.”
Intended Use and Contraindications
Intended Use
EndoVue series monitors are intended for use in a medical environment to display high quality video and graphic
images.
Contraindications
1. Do not use this product in the presence of flammable anesthetics mixture with air, oxygen or nitrous oxide.
2. No part of this product may come in contact with a patient. Never touch the product and a patient at the
same time.
3. This product is capable of displaying Radiology (PACS) images for reference, not diagnostic, purposes only.
4. For mission critical applications, we strongly recommend that a replacement unit be immediately available.
Model
BridgePower BM060S24F
AC Input
100 - 240 Volts, 50 to 60 Hz
DC Output
24 Volts, 2.7 Amps
Model
BridgePower BM060S24F
AC Input
100 - 240 Volts, 50 to 60 Hz
DC Output
24 Volts, 2.7 Amps
Summary of Contents for EndoVue 19"
Page 1: ...EndoVue USER MANUAL ENGLISH ...
Page 2: ......
Page 4: ......
Page 22: ...10 Connector Panel Overview ...
Page 26: ...14 Troubleshooting ...
Page 30: ...18 Electromagnetic Compatibility Tables ...
Page 31: ......