![NBN FLIR A6 Series User Manual Download Page 96](http://html1.mh-extra.com/html/nbn/flir-a6-series/flir-a6-series_user-manual_3670341096.webp)
Thermographic measurement techniques
3. Measure the radiation intensity (= apparent temperature) from the reflecting source
using the following settings:
• Emissivity: 1.0
• D
obj
: 0
You can measure the radiation intensity using one of the following two methods:
Figure 23.3
1 = Reflection source
Figure 23.4
1 = Reflection source
Using a thermocouple to measure reflected apparent temperature is not recommended
for two important reasons:
• A thermocouple does not measure radiation intensity
• A thermocouple requires a very good thermal contact to the surface, usually by gluing
and covering the sensor by a thermal isolator.
23.2.1.1.2
Method 2: Reflector method
Follow this procedure:
1. Crumble up a large piece of aluminum foil.
2. Uncrumble the aluminum foil and attach it to a piece of cardboard of the same size.
3. Put the piece of cardboard in front of the object you want to measure. Make sure that
the side with aluminum foil points to the camera.
4. Set the emissivity to 1.0.
#T559950; r. AD/35720/35720; en-US
88
Summary of Contents for FLIR A6 Series
Page 1: ...User s manual FLIR A6xx series nbn Austria GmbH...
Page 2: ......
Page 3: ...User s manual FLIR A6xx series T559950 r AD 35720 35720 en US iii...
Page 4: ......
Page 8: ......
Page 71: ...Mechanical drawings 16 T559950 r AD 35720 35720 en US 63...
Page 81: ...CE Declaration of conformity 17 T559950 r AD 35720 35720 en US 73...
Page 82: ......
Page 84: ...Digital I O connection diagrams 19 T559950 r AD 35720 35720 en US 76...
Page 85: ......
Page 125: ......