September 13, 2018
6-11
Electromagnetic Compatibility (EMC)
Compliance for not
losing stored patient
data
No change allowed. Check normal program start up before and after immunity
test are completed, to confirm test loads properly ensuring safe storing of data.
Compliance to not
burning
No fire or smoke are allowed during all immunity tests observe that the system is
free of fire or smoke from burned or overheated components.
The Essential Performance is verified after the tests
.