Specifications
Brainz BRM3 Brain Monitor Reference Manual |
36
EEG specifications
Artifact Rejection:
Automatic
Sensitivity:
±
500
μ
Vpk for maximum DAU output
Dynamic Range:
0.05 – 500
μ
Vpk (1-20 Hz)
Update Rate:
0.05 sec (EEG Waveform, Impedance)
1 minute (other parameters)
Event Markers:
User-selected
Impedance Checking:
Automatic
Check signals:
1.88 nArms @ 80, 84, 88, 92 and 96 Hz
aEEG Filter:
0-2 Hz:
60 dB/decade
2-12 Hz:
12 dB/decade
12-16 Hz:
1 dB > 10 Hz level
16-30 Hz:
-120 dB/decade
Spectral Edge:
90% for 2-20 Hz bandwidth
Computed parameters
Cross-cerebral and Bilateral amplitude-integrated EEG (aEEG) and Integrated Amplitude
EEG intensity
Impedance and AC Supply Noise
Spectral Edge Index (SEI).
Display modes
Real-time (EEG Waveform, aEEG, EEG Power Spectrum)
10 minute average (SEI, Integrated Amplitude)
Long term trend (SEI, Integrated Amplitude, Intensity, Impedance, AC Supply Noise)
Rapid numeric (Impedance).