Sensor
Sensor Impedance Number (k
Ù
) Sensor
Indicator
Left Front
0.7 - 1.3
Green
Left Back
1.8 - 2.6
Green
Right Front
4.0 - 5.4
Green
Right Back
7.5 - 9.2
Yellow
Figure 10: Normal display when test plug inserted
If the results described in the previous paragraph are not obtained:
1.
Follow any recommendations displayed on the
Signal Status
screen.
2.
If the problem persists, contact your healthcare institution’s service personnel, or contact an
Authorized Service Center. See
Servicing
on page 34.
Touch Cancel on the
Signal Status
screen to end the check for normal operation.