April 27, 2015
4-11
Nicolet EEGwireless32/64 Operation
Downloading data from the Amplifier to the PC
1. Locate the correct
Patient ID
in the Patient List.
2. Right-click
patient name > Acquire > NicoletOne
.
3. The Amplifier Contains Data dialogue box reappears, asking what you would like
to do with the Data on the Amplifier.
4. Select
Exit
and
Retrieve Data
.
NOTE:
If the Patient ID’s do not match, you will receive a message warning that the
ID’s are different and if you want to proceed.
The Recorder window will open and close briefly, and the File Catch-up Details box
will show the status of the exam being copied from the Amplifier to the PC. You may
minimize this window, however closing it will stop this process. File Catch-up will
need to complete before using the Amplifier again. Once this process completes, you
can choose to
Delete Data on Amplifier
when starting a new exam.
NOTE:
Choosing
Exit and Retrieve Data
in this fashion will download the EEG
data from the amplifier, and will not contain the annotations or events entered by the
technician at the beginning of the exam.
These annotations can be preserved by retrieving the data by reviewing the original
exam:
1. Connect the amplifier to system. Ensure it is not currently recording, press the
Stop
recording button on the amplifier if necessary.
2. Choose to review the original exam.
3. You will get a prompt stating that there is data on the amplifier and to
Download
the data now
, or to
Mark the exam as complete
.
4. Choose
Download the data now
. The Merge Process will start and append the
data to the end of this exam, including the original annotations/events.
Internal SpO2 operation
1. The amplifier supports the acquisition of SpO2 data. The amplifier has a built in
Nonin model for heart rate, pleth, SpO2. To acquire SpO2 data, connect a Nonin
finger probe with optional extension cables to the amplifier. Please refer to the
Nicolet Software Reference guide for instructions on using vital signs or an
external SpO2 device. See
Amplifier symbols and components
in Chapter 1 for
the connection port.
Summary of Contents for Nicolet EEGwireless32
Page 2: ......
Page 20: ...Nicolet Systems 1 2 April 27 2015 Blank page...
Page 32: ...Nicolet Systems 1 14 April 27 2015 Blank page...
Page 34: ...Nicolet Systems 2 2 April 27 2015 Blank page...
Page 40: ...Nicolet Systems 2 8 April 27 2015 Blank page...
Page 42: ...Nicolet Systems 3 2 April 27 2015 Blank page...
Page 52: ...Nicolet Systems 4 2 April 27 2015 Blank page...
Page 62: ...Nicolet Systems 4 12 April 27 2015 Blank page...
Page 64: ...Nicolet Systems 5 2 April 27 2015 Blank page...
Page 82: ...Nicolet Systems 5 20 April 27 2015...
Page 83: ...April 27 2015 6 1 6 Wireless Operation...
Page 84: ...Nicolet Systems 6 2 April 27 2015 Blank page...
Page 86: ...Nicolet Systems 6 4 April 27 2015 Blank page...
Page 88: ...Nicolet Systems 7 2 April 27 2015 Blank page...
Page 96: ...Nicolet Systems 8 2 April 27 2015 Blank page...
Page 102: ...Nicolet Systems 9 2 April 27 2015 Blank page...
Page 107: ...April 27 2015 10 1 10 Electromagnetic Compatibility EMC...
Page 108: ...Nicolet Systems 10 2 April 27 2015 Blank page...
Page 118: ...Nicolet Systems 10 12 April 27 2015 Blank page...
Page 120: ...Nicolet Systems 11 2 April 27 2015 Blank page...
Page 126: ...Nicolet Systems 11 8 April 27 2015 Blank page...
Page 128: ...Nicolet Systems 12 2 April 27 2015 Blank page...
Page 134: ...Nicolet Systems 12 8 April 27 2015 Blank page...