User and Service Manual
Natus® Brain Monitor &
Embla Dx Series™
43
Warnings and Cautions
WARNING:
Only use Natus approved equipment on the trolley/cart. Non-approved equipment
may compromise the function and safety of the system.
Make sure that any platform, table, cart, or other surface used during the operation, transport,
or temporary or permanent storage of the system and its components is adequate, sturdy, and
safe. Natus is not responsible for any injury or damage that may result from inadequate, poorly
constructed, or unapproved transports, carts, or operating surfaces. Natus is not responsible
for any injury or damage that may result from improper cable storage during transport.
WARNING:
Do not tilt the trolley/cart more than 10° incline as this will compromise the stability
of the trolley/cart.
TIPPING HAZARD:
During transport, the user should guide the cart using both hands,
ensuring the wheel base is aligned so that a single caster leads in the direction of motion.
Failure to lead the cart with one wheel could result in a tipping hazard when ascending or
descending steps or thresholds.
Electrical Input and Isolation Transformer Details
EU
Electrical input
200-240V AC, 2.24A @ 50 Hz
Isolation Transformer
Powervar ABC500-22MED
North America
Electrical input
120V AC, 3.10A @ 60 Hz
Isolation Transformer
Powervar ABC300-11MED
Japan
Electrical input
100V-120V AC, 8.0A, 50/60Hz
Isolation Transformer
Powertronix D1CSWFCNOE2