Potential Equalization Conductor
The Natus base unit provides a potential equalization conductor for optional use. To install, connect a
bus cable from the potential equalization conductor to the potential equalization busbar of the electrical
insulation in the room where the system is used.
Connection Mode
The following section describes how the Brain Monitor & Embla Dx Series Amplifiers system is
connected.
When the Natus base unit is physically connected to the breakout box using the patient cable, the
Brain
Monitor & Embla Dx Series amplifiers
are referred to as being in
Physical Connection Mode
. In this
mode, the data is transmitted over the wired connection to the base unit, which then transmits the data to
the computer.
The computer software will display and store this data.
Figure 1: Physical Connection of the System