7-38
Chapter 7
Operation
t
raiN
P
arameters
f
uNctioNs
To select
TRAIN PARAMETERS FUNCTIONS
proceed as follows.
1.
Press the
SETUP
touch-pad control.
2.
Toggle down the list to
TRAIN PARAMETERS FUNCTIONS
by using
the
TIMING
thumbwheel control.
3.
Press the
ENTER/ACCEPT
touch-pad control.
4.
S1 TRAIN DELAY, S2 TRAIN DELAY, TYPE OF TRAIN PARAMETERS,
and
TRAIN SYNC
LOCATION
will be displayed. Select the desired function by using the
TIMING
thumbwheel control.
5.
Press the
ENTER/ACCEPT
touch-pad control again.
S1 & S2 t
rain
D
elay
Applications occasionally require an adjustable delay before a train of
pulses is delivered to the preparation. The
S1 TRAIN DELAY
and
S2 TRAIN DELAY
function allows the placement of the delay prior to the onset of the train.
For example,
suppose it is desired to deliver a single train of pulses from
the S1 output with the following parameters:
Train Delay
= 1 second
Train Duration
= 200 ms
Number Pulses in Train = 10
Pulse Duration
= 1 ms
Output Amplitude
= 10 volts
The setup should be as follows:
1.
Press the
SETUP
touch-pad control and select
TRAIN PARAMETERS
FUNCTIONS
.
2.
Press
ENTER/ACCEPT
and select
S1 TRAIN DELAY
,
then
ENTER/ACCEPT
.
3.
In the next display, set the
S1 TRAIN DELAY
to 01.000 (seconds), and
press
ENTER/ACCEPT
.
4.
Select
S1
FUNCTION
and select
FUNCTION = SINGLE TRAIN (ST)
.
5.
Select the
TRAIN DURATION
and enter
00.200
(seconds).
6.
Select
TIMING
RATE
and enter
0050.00
PPS.
7.
Select
TIMING
DELAY
and set to minimum.
8.
Select
TIMING
DURATION
and set to
00.00100
(seconds).
9.
Set the
OUTPUT AMPLITUDE
to
010.00
(volts).
10. To initiate the train, press the
S1 SINGLE
touch-pad control.
Summary of Contents for Grass S88X
Page 2: ......
Page 10: ...x Table of Contents...
Page 12: ...1 12 Chapter 1 Specifications Glossary of Symbols Warnings Cautions...
Page 24: ...4 24 Chapter 4 Description of Circuits...
Page 54: ...8 54 Chapter 8 Artifacts Stimulus Isolation Constant Current...
Page 64: ...B 64 Chapter B Appendix...