S88X
Dual Output Square Pulse Stimulator
7-37
l
oad
s
etuP
& s
ave
s
etuP
When the POWER to the S88X is turned on, the S88X will display the last
set of parameters used. In addition, it is possible to save up to 9 sets of
stimulating parameters for use again.
S
etting
up
the
S88X p
arameterS
To set up the parameters to be used proceed as follows.
1.
Select the
FUNCTION, PULSES, RATE, DELAY, DURATION and OUTPUT AMPLITUDE
as is appropriate for the study.
b
aSic
m
oDeS
of
o
peration
later in this
chapter offers many basic setups.
Note:
To increase or decrease a value use the left and right arrow touch-
pad controls to select the digit to be changed. The number that is
flashing is the number that is selected. Then use the thumbwheel
control to increase or decrease that number.
S
aving
the
S88X S
etup
1.
Press the
SETUP
touch-pad control.
2.
Toggle down the list to
SAVE SETUP
by using
the
TIMING
thumbwheel
control and press the
ENTER/ACCEPT
touch-pad control.
3.
SAVE TO SETUP NUMBER x
will be displayed. (
NUMBER x
can be adjusted
from
1
to
9
by using the
TIMING
thumbwheel control.)
4.
Press the
ENTER/ACCEPT
touch-pad control again.
l
OaDing
a
S
et
OF
S88X S
aveD
p
arameterS
1.
Press the
SETUP
touch-pad control.
2.
Toggle down the list to
LOAD SETUP
by using
the
TIMING
thumbwheel
control and press the
ENTER/ACCEPT
touch-pad control.
3.
LOAD FROM SETUP NUMBER x
will be displayed. (
NUMBER x
can be adjusted
from
1
to
9
by using the
TIMING
thumbwheel control.)
4.
Press the
ENTER/ACCEPT
touch-pad control again.
Summary of Contents for Grass S88X
Page 2: ......
Page 10: ...x Table of Contents...
Page 12: ...1 12 Chapter 1 Specifications Glossary of Symbols Warnings Cautions...
Page 24: ...4 24 Chapter 4 Description of Circuits...
Page 54: ...8 54 Chapter 8 Artifacts Stimulus Isolation Constant Current...
Page 64: ...B 64 Chapter B Appendix...