Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems within Professional Healthcare use environment
Aurical Aud is intended for use in the electromagnetic environment specified below. The user of Aurical Aud should ensure that it is used in such an envir-
onment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz to 80 MHz
6 V rms
ISM Bands and Amateur
3 V rms
150 kHz to 80 MHz
6 V rms
ISM Bands and Amateur
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
Proximity fields from RF wire-
less communications
IEC 61000-4-3
27 V/m
385 MHz
28 V/m
450 MHz
9 V/m
710 MHz, 745 MHz, 780 MHz
28 V/m
810 MHz, 870 MHz, 930 MHz
28 V/m
1720 MHz, 1845 MHz, 1970 MHz
28 V/m
2450 MHz
9 V/m
5240 MHz, 5500 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz
9 V/m
710 MHz, 745 MHz, 780 MHz
28 V/m
810 MHz, 870 MHz, 930 MHz
28 V/m
1720 MHz, 1845 MHz, 1970 MHz
28 V/m
2450 MHz
9 V/m
5240 MHz, 5500 MHz, 5785 MHz
Separation distance between any electronic parts of Aur-
ical Aud and any RF wireless communication equipment
must be more than 30 cm (11.8 inches).
Note
: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
IEC 60601-1-2:2007 and EN 60601-1-2:2007
Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems
Aurical Aud is intended for use in the electromagnetic environment specified below. The user of Aurical Aud should ensure that it is used in such an envir-
onment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR11
Group 1
Aurical Aud uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR11
Class B
Aurical Aud is suitable for use in all environments, including domestic environments and those
directly connected to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
36
Aurical Aud
17 Technical specifications