MAINTENANCE 

 

WARNING:  For your own safety, turn switch off and remove plug from power source before 
maintaining your microscope.  If the power cord is worn, cut or damaged in any way, have it 
replaced immediately to avoid shock or fire hazard. 

 
1.   OPTICAL MAINTENANCE

 

 

A.  Do not attempt to disassemble any lens components.  Consult a microscope service technician 

when any repairs not covered by instructions are needed. 

 
B.  Prior to cleaning any lens surface, brush dust or dirt off lens surfaces using a camel hair brush.  

Or use air to blow dust and lint off surfaces.  Use of compressed air in a can,  available at any 
computer supply store, is a good source of clean air. 

 

C.  To clean eyepiece lenses, do not remove from eyepiece tube.  Clean only the outer lens surface.  

Breath on lens to dampen surface, then wipe with lens paper or tissue or use a cotton swab 
moistened with distilled water.  Wipe lenses with a circular motion, applying as little pressure as 
possible.  Avoid wiping dry lens surface as lenses are scratched easily.  If excessive dirt or grease 
gets on lens surfaces, a small amount of Windex can be used on a cotton swab or lens tissue.  To 
clean objective lenses.  Clean front lens element only, following same procedure. 

 
 
2.  MECHANICAL MAINTENANCE 
 

The only mechanical adjustment you might ever require is the tension of the focusing mechanism.  
This has been adjusted at the factory, but over the course of time it may loosen and cause the 
head of the microscope to slip downward on the focusing block. 
 
The knurled 6mm wide by 35mm diameter tension adjustment collar is located between arm and 
coarse focus knob on right side of microscope.  With the supplied “L” hex wrench, loosen the set 
screw located in only one of the four holes on tension adjustment collar.  Using supplied rod 
wrench turn tension adjustment collar clockwise to tighten tension, counterclockwise to loosen 
tension.  After adjusting tighten the set screw to lock collar in place. 

 

NOTE:  It is recommended that you leave the tension as loose as possible for ease of focusing, 
yet not so loose that it permits the head of microscope to drift downward from its own weight and 
cause the microscope to “drift” out of focus. 

 
3.  ELECTRICAL MAINTENANCE 
 

The extent of electrical maintenance, by other than a qualified technician, should be bulb replacement.  
BE CERTAIN TO TURN SWITCHES OFF AND REMOVE PLUG FROM POWER SOURCE OUTLET 
BEFORE CHANGING BULBS. 

 

A.  To replace top bulb (National bulb #800-423)....remove light shade by rotating in a counter-

clockwise direction.  Remove light bulb by firmly grasping edge of reflector and pulling straight out 
from bi-pin socket.  Note that the socket holds bulb securely, so you might have to pull rather 
firmly.  Using a cloth, hold new reflector bulb and gently push new bulb in bi-pin socket.  Replace 
light shade. 

 
B.  To replace bottom bulb in stand (National bulb #800-138)....remove perforated lamp cover plate 

located on bottom of base by removing the large black knurled locking screw then push cover 
plate to the rear of base.  It will be necessary to cut the white cable tie wrapped around base of 
lamp (this tie is used to secure lamp during shipment and it is not necessary to replace).  Holding 
fluorescent lamp with a cloth, gently pull lamp straight out from socket.  Push new lamp into place 
in same manner and replace cover plate and locking screw. 

 

Summary of Contents for 460TBL-10

Page 1: ...ientific Instruments Inc 6508 Tri County Parkway Schertz Texas 78154 Phone 210 590 9010 Fax 210 590 1104 INSTRUCTIONS FOR 460TBL 10 STEREOSCOPIC ZOOM MICROSCOPES Copyright 1 2 01 National Optical Scientific Instrument Inc ...

Page 2: ...Shade Main Power Switch Stage Clip Stage Plate translucent white illustrated Rubber eyepiece shields Eyepiece locking set screw Tension adjustment collar located right side between coarse focus knob and arm Focusing knob both sides Light directional lever Arm Fuse Incidental top light switch Transmitted bottom light switch Stage plate locking set screw ...

Page 3: ...otection Use of the styrofoam alone will not provide adequate protection in transit and will void your warranty ASSEMBLY OPERATION 1 Install rubber eyepiece shields over top of eyepieces with the flared portion of the shields positioned at the outside edge of eyepieces These help block out undesired light reflections and to position your eyes at the proper point above the eyepieces 2 Please observ...

Page 4: ...ight diopters in a clockwise direction to the lowest position E Rotate zoom control turret to the highest magnification by aligning the number 4 on magnification number ring to the white index dot on zoom control turret F While looking through right eyepiece with one eye rotate focusing control knob until specimen comes into sharp focus through right eyepiece G Rotate zoom control turret to the lo...

Page 5: ...is located between arm and coarse focus knob on right side of microscope With the supplied L hex wrench loosen the set screw located in only one of the four holes on tension adjustment collar Using supplied rod wrench turn tension adjustment collar clockwise to tighten tension counterclockwise to loosen tension After adjusting tighten the set screw to lock collar in place NOTE It is recommended th...

Page 6: ...nsion control Poor resolution image not sharp Objective lenses dirty Clean objective lenses Eyepiece lens dirty Clean eyepiece lenses Spots in field of view Eyepiece lens dirty Clean eyepiece lenses Spots in field of view can also result from dirt on inside of eyepiece It is recommended that you have service technician clean inside of lens OPTIONAL ACCESSORIES AND PARTS 615 400 WF15X Eyepieces pai...