This model is furnished with an adjustment for the incidental (top) light.  To adjust, move light 
directional lever up or down to position the beam of light providing optimum illumination to specimen. 

 
6.  INTERPUPILLARY ADJUSTMENT  
 

This permits each user to adjust spacing between eyepieces in order to accommodate distance 
between their eyes.  While looking through the microscope eyepieces with both eyes, grasp eyepiece 
tube housings with both hands and rotate them on their axis, moving eyepieces apart or together until 
a full field of view is observed and images blend into one.  Interpupillary distance is now corrected for 
your own inter-ocular distance and does not require further adjustment later unless another user 
changes this adjustment. 

 
7.  FOCUSING 
 

A.  Rotate zoom control turret to the lowest magnification by aligning the number “1” on magnification 

number ring to the white index dot on zoom control turret.  Lower magnifications have larger fields 
of view, making it easier to position and locate area to be viewed. 

 
B.  Place a flat object or specimen slide (cover glass up), on center of stage plate. 
 
C.  Turn focusing knobs until object being viewed is in focus. 
 
D.  Both eyepieces have knurled diopter adjustment rings.  Rotate both left and right diopters in a 

clockwise direction to the lowest position. 

 
E.  Rotate zoom control turret to the highest magnification by aligning the number “4” on 

magnification number ring to the white index dot on zoom control turret. 

 
F.  While looking through right eyepiece with one eye, rotate focusing control knob until specimen 

comes into sharp focus through right eyepiece. 

 
G.  Rotate zoom control turret to the lowest magnification. 
 
H.  Adjust the right diopter until the image is sharp.  Do not change the focusing knob position. 
 
I. 

Without changing the position of the focusing knob, adjust the left eyepiece diopter until you 
obtain a sharp image in left eyepiece.  the image should now be sharp throughout the zoom power 
range. 

 
8.  For optimum sharpness of image in the entire field of view, it is important to position your eyes at the 

correct point above the eyepieces.  Looking through both eyepieces, slowly move your eyes towards 
eyepieces to a level where clarity of the entire field of view is achieved.  After a brief period of viewing, 
you will easily find the best point for your vision. 

 

Specification Chart 

 

Zoom 

Objective 

Position 

Eyepiece 

 

WF10X 

Field No. 20 

(supplied) 

(accepts reticle 23mm O.D.)

 

WF15X 

Field No. 13 

(optional)

 

Working 

Distance 

Maximum 

Specimen 

Height 

Total 

Magnification 

Field 

Size 

Total 

Magnification 

Field 

Size 

1X 

10X 

20mm 

15X 

13mm 

70mm 

36mm 

1.5X 

15X 

13.3mm 

22.5X 

8.7mm 

2X 

20X 

10mm 

30X 

6.5mm 

2.5X 

25X 

8mm 

37.5X 

5.2mm 

3X 

30X 

6.7mm 

45X 

4.3mm 

Summary of Contents for 460TBL-10

Page 1: ...ientific Instruments Inc 6508 Tri County Parkway Schertz Texas 78154 Phone 210 590 9010 Fax 210 590 1104 INSTRUCTIONS FOR 460TBL 10 STEREOSCOPIC ZOOM MICROSCOPES Copyright 1 2 01 National Optical Scientific Instrument Inc ...

Page 2: ...Shade Main Power Switch Stage Clip Stage Plate translucent white illustrated Rubber eyepiece shields Eyepiece locking set screw Tension adjustment collar located right side between coarse focus knob and arm Focusing knob both sides Light directional lever Arm Fuse Incidental top light switch Transmitted bottom light switch Stage plate locking set screw ...

Page 3: ...otection Use of the styrofoam alone will not provide adequate protection in transit and will void your warranty ASSEMBLY OPERATION 1 Install rubber eyepiece shields over top of eyepieces with the flared portion of the shields positioned at the outside edge of eyepieces These help block out undesired light reflections and to position your eyes at the proper point above the eyepieces 2 Please observ...

Page 4: ...ight diopters in a clockwise direction to the lowest position E Rotate zoom control turret to the highest magnification by aligning the number 4 on magnification number ring to the white index dot on zoom control turret F While looking through right eyepiece with one eye rotate focusing control knob until specimen comes into sharp focus through right eyepiece G Rotate zoom control turret to the lo...

Page 5: ...is located between arm and coarse focus knob on right side of microscope With the supplied L hex wrench loosen the set screw located in only one of the four holes on tension adjustment collar Using supplied rod wrench turn tension adjustment collar clockwise to tighten tension counterclockwise to loosen tension After adjusting tighten the set screw to lock collar in place NOTE It is recommended th...

Page 6: ...nsion control Poor resolution image not sharp Objective lenses dirty Clean objective lenses Eyepiece lens dirty Clean eyepiece lenses Spots in field of view Eyepiece lens dirty Clean eyepiece lenses Spots in field of view can also result from dirt on inside of eyepiece It is recommended that you have service technician clean inside of lens OPTIONAL ACCESSORIES AND PARTS 615 400 WF15X Eyepieces pai...