Configuring the NI 5684 in MAX

Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX
informs other programs about which devices reside in the system and how they are configured.
MAX is automatically installed with NI-568x.
1.

Launch MAX.

2.

In the configuration tree, expand 

Devices and Interfaces

 to see the list of installed

devices.

Note

 

If you do not see your device listed, press <F5> to refresh the list of

installed devices. If the device is still not listed, power off the system, ensure
the device is correctly installed, and restart.

3.

Record the device identifier MAX assigns to the hardware. Use this identifier when
programming the NI 5684.

4.

Self-test the device by selecting the device in the configuration tree and clicking 

Self-Test

in the MAX toolbar.

The MAX self-test performs a basic verification of hardware resources.

NI RF Power Meter Programming Options

You can acquire data with the NI 5684 by using the NI-568x soft front panel (SFP), or you can
control the NI RF power meter programmatically using the NI-568x instrument driver.

8

  | 

ni.com

  | 

NI 5684 Getting Started Guide

Summary of Contents for USB-5684

Page 1: ...4 if you use it in a manner not described in this document Contents Electromagnetic Compatibility Guidelines 1 Verifying the System Requirements 2 Unpacking the Kit 2 Verifying the Kit Contents 3 Prep...

Page 2: ...ft Caution To ensure the specified EMC performance the length of the USB cable must be no longer than 30 m 100 ft Verifying the System Requirements To use the NI 5684 your system must meet certain re...

Page 3: ...ontents you need the following items 1 N m torque wrench Desktop computer laptop or chassis with one available USB 2 0 or USB 1 1 compliant slot Preparing the Environment Ensure that the environment y...

Page 4: ...see access and security messages during installation Accept the prompts to complete the installation 4 When the installer completes select Restart in the dialog box that prompts you to restart shut d...

Page 5: ...ay need to use the included Type N to SMA adapter depending on the RF source connection Caution Always refer to the specifications document before connecting signals Failure to observe the specified m...

Page 6: ...l 3 1 2 4 1 Status LED 2 External Trigger In Terminal 3 USB 2 0 A to Mini B Port 4 External Trigger Out Terminal Note External Trigger Out capability is currently unsupported Connector Care and Handli...

Page 7: ...ermanently damages the internal components Before making connections verify the maximum input levels labeled below the ports on the module front panel Avoid Mechanical Shock Precision connectors are d...

Page 8: ...e device is still not listed power off the system ensure the device is correctly installed and restart 3 Record the device identifier MAX assigns to the hardware Use this identifier when programming t...

Page 9: ...n modify an NI 568x C example to create an application with Microsoft Visual C C and all required include and library files are automatically added to the project C or LabWindows CVI Available at Prog...

Page 10: ...pport ni com support RF Solutions ni com rf Updates ni com updates Services ni com services Getting Started with LabVIEW Getting Started with LabWindows CVI Located using the NI Example Finder Located...

Page 11: ...oC for your product by visiting ni com certification If your product supports calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corpora...

Page 12: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance p...

Reviews: