National Instruments TB-9212 Calibration Procedure Download Page 8

8

|

ni.com

|

TB-9212 Calibration Procedure

Notes

Make sure the thermocouple calibrator and NI 9212 are completely settled 

at the desired temperature setting before acquiring data. 

Wait at least 300 ms for the Ectron 1140A to settle before acquiring data with the 
NI 9212.

Wait at least 1 second to compensate for the open thermocouple settling time of the 
NI 9212.

d.

Average the readings for each channel. Record the average of the readings for each as 
Temp

ai

x

 

where 

x

 is the channel number.

e.

Clear the task.

10. Set the thermocouple calibrator output to CJC

temp1

 

to obtain the readings for the next four 

physical channels, ai4:7. This ensures that during the thermocouple measurements, the 
input channel measures very close to 0 V, which eliminates gain errors in the NI 9212.

11. Repeat step 9.
12. Perform the following calculation using the recorded Temp

ai

x

 and CJC

temp

 

x

 values.

Temp

ErrorCh0

:3

 = 

Temp

ai0:3

 - 

CJC

temp0

Temp

ErrorCh4:7

 = 

Temp

ai4:7

 - 

CJC

temp1

13. Compare the calculation result for each channel to the Upper Limit and Lower Limit values 

in Table 6.

14. Remove the TB-9212 from the NI 9212.
15. Disconnect the barrier strip assembly from the TB-9212.

Note

The test limits listed in Table are based upon the 374389A-01 edition of the 

NI 9212 with NI TB-9212 Datasheet.

 Refer to the most recent 

NI 9212 with 

NI TB-9212 Datasheet

 online at 

ni.com/manuals

If the verification procedure determines that the TB-9212 is outside of the limits, refer to 

Where 

to Go for Support

 for assistance in returning the device to NI.

Table 6.  

TB-9212 Verification Test Limits

Connector 

Timing 

Mode

Test Point 

Value (°C)

Error

1-Year Limits (°C)

Lower 

Limit

Upper 

Limit

TB-9212 

with Screw 

Terminal

High 

Resolution

CJC

temp

x

Temp

ErrorCh

x

-0.44

0.44

TB-9212 

with 

Mini-TC

-1.1

1.1

Summary of Contents for TB-9212

Page 1: ...sively referred to as TB 9212 For more information about calibration solutions visit ni com calibration Note Refer to the NI 9212 Calibration Procedure on ni com manuals for instructions on calibratin...

Page 2: ...com and help files install with the software NI cDAQ 9174 9178 Quick Start NI DAQmx installation and hardware setup NI 9212 Getting Started Guide NI 9212 with TB 9212 specific information NI 9212 Dat...

Page 3: ...212 An NI 9212 module that is verified to be within product specifications Chassis cDAQ 9178 Barrier Strip Assembly Thermocouple Wire TT E 24 SLE 100 E type SLE 24 AWG 20 m to 30 m Mini Thermocouple C...

Page 4: ...re assembly Crimp the chromel spade lugs to the chromel wire to create 8 positive spade lugs Crimp the constantan spade lugs to the constantan wire to create 8 negative spade lugs 2 Use the thermocoup...

Page 5: ...how to install the software and hardware and how to configure the device in Measurement Automation Explorer MAX Complete the following steps to set up the TB 9212 1 Install NI DAQmx Note Ensure that...

Page 6: ...ou are using a TB 9212 with mini TC you do not need to remove the top cover 2 Connect the thermocouple calibrator to the TB 9212 using the barrier strip assembly created in the Building the Barrier St...

Page 7: ...eliminates gain errors in the NI 9212 9 Acquire and average samples a Create and configure an AI temperature thermocouple channel for the NI 9212 according to Table 4 b Configure the AI temperature th...

Page 8: ...Repeat step 9 12 Perform the following calculation using the recorded Tempaix and CJCtemp x values TempErrorCh0 3 Tempai0 3 CJCtemp0 TempErrorCh4 7 Tempai4 7 CJCtemp1 13 Compare the calculation resul...

Page 9: ...complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Appli...

Reviews: