TB-9212 Calibration Procedure

Software

Calibrating the TB-9212 requires the installation of NI-DAQmx 15.1 or later on the calibration 
system. You can download NI-DAQmx from 

ni.com/downloads

. NI-DAQmx supports 

LabVIEW, LabWindows

/CVI

, ANSI C, and .NET. When you install NI-DAQmx, you only 

need to install support for the application software that you intend to use.

Documentation

Consult the following documents for information about the TB-9212, NI-DAQmx, and your 
application software. All documents are available on 

ni.com

 and help files install with the 

software.

NI cDAQ-9174/9178 Quick Start

NI-DAQmx installation and hardware setup

NI 9212 Getting Started Guide

NI 9212 with TB-9212 specific information

NI 9212 Datasheet

NI 9212 with TB-9212 specifications and calibration interval

NI-DAQmx Readme

Operating system and application software support in NI-DAQmx

LabVIEW Help

LabVIEW programming concepts and reference information about NI-DAQmx 
VIs and functions

NI-DAQmx C Reference Help

Reference information for NI-DAQmx C functions and NI-DAQmx C properties

NI-DAQmx .NET Help Support for Visual Studio

Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET 
properties, key concepts, and a C enum to .NET enum mapping table

Summary of Contents for TB-9212

Page 1: ...sively referred to as TB 9212 For more information about calibration solutions visit ni com calibration Note Refer to the NI 9212 Calibration Procedure on ni com manuals for instructions on calibratin...

Page 2: ...com and help files install with the software NI cDAQ 9174 9178 Quick Start NI DAQmx installation and hardware setup NI 9212 Getting Started Guide NI 9212 with TB 9212 specific information NI 9212 Dat...

Page 3: ...212 An NI 9212 module that is verified to be within product specifications Chassis cDAQ 9178 Barrier Strip Assembly Thermocouple Wire TT E 24 SLE 100 E type SLE 24 AWG 20 m to 30 m Mini Thermocouple C...

Page 4: ...re assembly Crimp the chromel spade lugs to the chromel wire to create 8 positive spade lugs Crimp the constantan spade lugs to the constantan wire to create 8 negative spade lugs 2 Use the thermocoup...

Page 5: ...how to install the software and hardware and how to configure the device in Measurement Automation Explorer MAX Complete the following steps to set up the TB 9212 1 Install NI DAQmx Note Ensure that...

Page 6: ...ou are using a TB 9212 with mini TC you do not need to remove the top cover 2 Connect the thermocouple calibrator to the TB 9212 using the barrier strip assembly created in the Building the Barrier St...

Page 7: ...eliminates gain errors in the NI 9212 9 Acquire and average samples a Create and configure an AI temperature thermocouple channel for the NI 9212 according to Table 4 b Configure the AI temperature th...

Page 8: ...Repeat step 9 12 Perform the following calculation using the recorded Tempaix and CJCtemp x values TempErrorCh0 3 Tempai0 3 CJCtemp0 TempErrorCh4 7 Tempai4 7 CJCtemp1 13 Compare the calculation resul...

Page 9: ...complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Appli...

Reviews: