Test Equipment
Table 1 lists the equipment recommended for the performance verification procedure. If the
recommended equipment is not available, select a substitute using the requirements listed in
Table 1.
Table 1.
Recommended Equipment
Equipment
Recommended
Model
Requirements
Thermocouple Calibrator
Ectron 1140A
A high-precision
thermocouple source with an
accuracy of
≤
0.11 °C and an
output impedance of
≤
10
Ω
.
NI 9212
—
An NI 9212 module that is
verified to be within product
specifications.
Chassis
cDAQ-9178
—
Barrier Strip
Assembly
Thermocouple
Wire
TT-E-24-SLE-100
E-type SLE 24 AWG, 20 m to
30 m
Mini
Thermocouple
Connector
Omega
SMPW-E-M
E-type, must fit the output
terminal of the thermocouple
calibrator
Thermocouple
Barrier Strips
Omega BS16A
8 channels
Thermocouple
Barrier Strip
Jackets (x8)
Omega BSJ-E
E-type, 8 channels
Thermocouple
Spade Lugs (x32)
Omega SLCH-20
16 lugs of chromel to match
the E-type wires
Omega SLCO-20
16 lugs of constantan to match
the E-type wires
Spade-Lug
Crimping Tool
Omega
CRIMPING
TOOL-P
Must effectively crimp the
spade lugs with the 24 AWG
wire.