background image

 

National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the 

Terms of Use

 section on 

ni.com/legal

 for more information about National Instruments trademarks. Other product and company names mentioned herein are 

trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate 
location: 

Help»Patents

 in your software, the 

patents.txt

 file on your media, or the 

National Instruments Patent Notice

 at 

ni.com/patents

.

© 2004–2008 National Instruments Corporation. All rights reserved.

373959D

Sep08

Specifications

This section lists additional specifications for the NI TB-2641 when used 
with the NI PXI-2532. All specifications are subject to change without 
notice. Visit 

ni.com/manuals

 for the most current specifications.

Input Characteristics

All input characteristics are DC, AC

pk

, or a combination unless otherwise 

specified.

Maximum switching voltage

Channel-to-channel..........................100 V

Channel-to-ground...........................100 V, CAT I

Caution

This module is rated for Measurement Category I and intended to carry 

signal voltages no greater than 100 V. This module can withstand up to 500 V impulse 
voltage. Do

not

 use this module for connections to signals or for measurements within 

Categories II, III, or IV. Do 

not

 connect to MAINs supply circuits (for example, wall 

outlets) of 115 or 230 VAC. Refer to the 

Read Me First: Safety and Electromagnetic 

Compatibility

 document for more information on measurement categories.

When hazardous voltages (>42.4 V

pk

/60 VDC) are present on any relay terminal, safety 

low-voltage (<42.4 V

pk

/60 VDC) cannot be connected to any other relay terminal.

Maximum current (per channel) .............0.5 A

DC path resistance ..................................<1.5 

Ω

Row and column protection
resistors (when used)  .............................100 

Ω

Bandwidth (–3 dB, 50 

Ω

 termination)

Typical

.............................................≥

9 MHz

Crosstalk (50 

Ω

 termination)

Channel-to-channel

10 kHz ......................................<–75 dB

100 kHz ....................................<–60 dB

1 MHz.......................................<–40 dB

Summary of Contents for TB-2641

Page 1: ...ll the terminal block Make sure you have the following NI TB 2641 terminal block 1 8 in flathead and 1 Phillips screwdrivers Eight 2 mm jumpers Two 34 conductor 28 AWG 050 in pitch ribbon cable assemblies not included One 16 conductor 28 AWG 050 in pitch ribbon cable assembly not included Note Refer to the Accessories section for information about ordering the appropriate cable assemblies Connecti...

Page 2: ...he terminal block housing 6 Connect each ribbon cable to the appropriate headers on the column connection board and the module interface board 7 Reassemble the terminal block Note For information about protection resistance and matrix expansion refer to the NI Switches Help Figure 1 NI TB 2641 Terminal Block 1 Top Cover Screw 2 Top Cover 3 Strain relief Screws 4 Strain relief Bar 5 Column Connecti...

Page 3: ...ccessories for the NI TB 2641 Accessory Manufacturer Part Number NI SCB 264X screw terminal block National Instruments 779341 01 Row and column cable kit for the NI SCB 264X terminal blocks National Instruments 779346 01 Row connection cable for column expansion National Instruments 779325 01 16 Conductor 05 in pitch ribbon cable assemblies Samtec FFSD 08 01 N 34 Conductor 05 in pitch ribbon cable...

Page 4: ...eristics All input characteristics are DC ACpk or a combination unless otherwise specified Maximum switching voltage Channel to channel 100 V Channel to ground 100 V CAT I Caution This module is rated for Measurement Category I and intended to carry signal voltages no greater than 100 V This module can withstand up to 500 V impulse voltage Do not use this module for connections to signals or for m...

Reviews: