SCXI-1520 Calibration Procedure
Calibration
The calibration process consists of the following steps:
1.
Setting up the device for testing.
2.
Verifying the existing operation of the device to determine whether
it is operating within its specifications.
3.
Adjusting the device with respect to a known voltage source.
4.
Verifying that the device is operating within its specifications after
adjustments.
Setting Up the SCXI-1520 for Testing
Complete the following steps to set up the SCXI-1520 for testing.
1.
Install the SCXI-1520 in slot 1 of the SCXI chassis.
2.
Using an SCXI-1349 cable adapter, connect a 68-to-68-pin cable
between the SCXI-1520 module and the DAQ device installed in the
host computer, as shown in Figure 1.
Figure 1.
Connecting an SCXI Chassis to a DAQ Device
1
DAQ Device
2
SH68-68 Cable
3
SCXI-1349 Cable Adapter
4
Thumbscrews
5
SCXI-1520 Module
6
SCXI-1314 Terminal Block
7
SCXI Chassis Power Switch
8
SCXI-1000 Chassis
®