![background image](http://html1.mh-extra.com/html/national-instruments/scxi-1520/scxi-1520_calibration-procedure_3577183021.webp)
©
National Instruments Corporation
21
SCXI-1520 Calibration Procedure
–0.158060
31
–0.158352
–0.157777
25
0.136110
36
0.135864
0.136358
26
0.000000
36
–0.000111
0.000111
26
–0.136110
36
–0.136358
–0.135864
26
0.116670
42
0.116455
0.116879
27
0.000000
42
–0.000095
0.000095
27
–0.116670
42
–0.116879
–0.116455
27
0.087500
56
0.087341
0.087659
28
0.000000
56
–0.000071
0.000071
28
–0.087500
56
–0.087659
–0.087341
28
0.075380
65
0.075248
0.075521
29
0.000000
65
–0.000062
0.000062
29
–0.075380
65
–0.075522
–0.075248
29
0.065330
75
0.065215
0.065452
30
0.000000
75
–0.000053
0.000053
30
–0.065330
75
–0.065452
–0.065215
30
0.056320
87
0.056220
0.056424
31
0.000000
87
–0.000046
0.000046
31
–0.056320
87
–0.056424
–0.056220
31
0.049000
100
0.048891
0.049109
32
0.000000
100
–0.000060
0.000060
32
–0.049000
100
–0.049109
–0.048891
32
0.042610
115
0.042507
0.042710
33
0.000000
115
–0.000059
0.000059
33
–0.042610
115
–0.042710
–0.042507
33
0.037690
130
0.037597
0.037788
34
0.000000
130
–0.000058
0.000058
34
Table 2.
SCXI-1520 Gain and Offset Calibration Table (Continued)
Test Point (V)
Gain
Lower Limit (V)
Upper Limit (V)
Index