National Instruments Corporation
19
SCXI-1520 Calibration Procedure
–2.041670
2.4
–2.045375
–2.037958
7
1.814810
2.7
1.811519
1.818111
8
0.000000
2.7
–0.001481
0.001481
8
–1.814810
2.7
–1.818111
–1.811519
8
1.580650
3.1
1.577774
1.583516
9
0.000000
3.1
–0.001290
0.001290
9
–1.580650
3.1
–1.583516
–1.577774
9
1.361110
3.6
1.358639
1.363583
10
0.000000
3.6
–0.001111
0.001111
10
–1.361110
3.6
–1.363583
–1.358639
10
1.166670
4.2
1.164548
1.168786
11
0.000000
4.2
–0.000952
0.000952
11
–1.166670
4.2
–1.168786
–1.164548
11
0.875000
5.6
0.873411
0.876589
12
0.000000
5.6
–0.000714
0.000714
12
–0.875000
5.6
–0.876589
–0.873411
12
0.753850
6.5
0.752477
0.755215
13
0.000000
6.5
–0.000615
0.000615
13
–0.753850
6.5
–0.755215
–0.752477
13
0.653330
7.5
0.652147
0.654520
14
0.000000
7.5
–0.000533
0.000533
14
–0.653330
7.5
–0.654520
–0.652147
14
0.563220
8.7
0.562195
0.564241
15
0.000000
8.7
–0.000460
0.000460
15
–0.563220
8.7
–0.564241
–0.562195
15
0.490000
10
0.489110
0.490890
16
0.000000
10
–0.000400
0.000400
16
Table 2.
SCXI-1520 Gain and Offset Calibration Table (Continued)
Test Point (V)
Gain
Lower Limit (V)
Upper Limit (V)
Index