background image

©

 National Instruments Corporation

5

SCXI-1314 Universal Strain Terminal Block

Installing the Terminal Block

To connect the terminal block to the SCXI module front connector, perform 
the following steps:

1.

Connect the module front connector to its mating connector on the 
terminal block.

2.

Tighten the top and bottom thumbscrews on the back of the terminal 
block to hold it securely in place.

Cleaning the Terminal Block

Clean the terminal block by brushing off light dust with a soft, nonmetallic 
brush. Remove other contaminants with deionized water and a stiff 
nonmetallic brush. The unit must be completely dry and free from 
contaminants before returning it to service.

Calibrating the Terminal Block

For information on calibrating the SCXI-1314 see the 

SCXI-1314 

Universal Strain Terminal Block Calibration Procedure

 located online at 

ni.com/calibration

 

and follow the link for 

Manual Calibration 

Procedures

.

Specifications

All specifications are typical at 25 °C unless otherwise specified.

Electrical

Accuracy of resistors

120

Ω 

completion resistor

............... ±

0.1%

350

 completion resistor

............... ±

0.1%

100 k

 shunt calibration resistor

.... ±

0.1%

Temperature coefficient of resistors

120

Ω 

completion resistor

............... ±

10 ppm/°C

350

 completion resistor

............... ±

10 ppm/°C

100 k

 shunt calibration resistor

.... ±

10 ppm/°C