RM-26999 Calibration Procedure
© National Instruments Corporation
5
Recommended Documentation
Go to
to locate the following documentation for more information
when performing this calibration:
RM-26999 Specifications
RM-26999 Getting Started Guide
RM-26999 User Manual
Test Equipment
This section details the equipment NI recommends for each test performed as part
of this calibration procedure.
NI Calibration Executive Users
Refer to the Calibration Executive Help to find an updated list of test
equipment for this calibration procedure.
Standard
Recommended Model
Where Used
Functional Requirement(s)
Calibrator
Fluke 5730A
Verifications:
DC Gain
AC Gain
DC Range:
±
1000 V
DC Accuracy:
±
20 µV/V
AC Range:
±
1000 V
AC Accuracy:
< 100 kHz:
±
0.3%
< 1 MHz:
±
1.5%
Calibrator Amp
Fluke 5725A
Verifications:
AC Gain
AC Range:
±
1000 V
AC Accuracy:
< 100 kHz:
±
0.3%
< 1 MHz:
±
1.5%