RM-26999 Calibration Procedure
© National Instruments Corporation
17
Perform Adjustment
Perform an adjustment at least once within the calibration interval. Adjustment
automatically updates the calibration constants, the date, and the temperature in
the DUT EEPROM. If the DUT passes the verification procedures within the As-Left
test limits, an adjustment is not required.
Note
NI recommends an adjustment to optimize the measurement performance
of the DUT even if it passes verification within the As-Left limits.
Initial Test Connection
Caution: Possibility of Electric Shock
Power off all high-voltage signals before making any connections.