background image

RM-26999 Calibration Procedure 

2

 ni.com/calibration 

Terms and Definitions 

DUT 

DUT is an acronym for Device Under Test, and refers to the NI product being 

calibrated. For this procedure, DUT refers to the RM-26999.

 

 

As-Found Limits 

These limits are derived from the published specifications for the DUT. NI 

uses these limits to determine if the DUT is performing within the 

recommended calibration interval specifications at the time of calibration 

and before any adjustment is performed. 

As-Left Limits 

These limits are derived from the published specifications for the DUT minus 

guardband to ensure a high probability that the DUT will meet its 

specifications over the next recommended calibration interval. 

Functional Test  

Functional Tests determine whether the DUT is operating correctly. 

Functional tests are not directly related to performance specifications. 

Verification 

Verification evaluates the measured calibration results against the defined 

As-Found Limits. The result of the evaluation is expressed as a Pass/Fail 

condition in the calibration certificate using an established evaluation 

formula.  

Adjustment 

Adjustment performs a set of operations on the DUT to optimize the 

measurement performance and conform it to the assigned calibrated values. 

Reverification 

Reverification evaluates the measured calibration results against the As-Left 

limits after adjustment. The As-Left limits may be tighter than the As-Found 

limits. 

Recommended 

Calibration 

Interval 

This interval indicates the recommended period between each round of 

verification and adjustment of the DUT. There is a high probability that, 

within this interval, the DUT will remain within the published warranted 

performance specifications. Some measurement DUTs have warranted 

specifications for different calibration intervals, for example: 24 hours, 90 

days, 1 year, and 2 years. In this case, the specification depends on the 

calibration cycle chosen by the user. 

Calibration Overview 

Recommended Calibration Interval

 

1 year 

 

Summary of Contents for RM-26999

Page 1: ...December 2020 RM 26999 Calibration Procedure...

Page 2: ...re procedure before beginning the calibration process Contents Terms and Definitions 2 Calibration Overview 2 Calibration Condition Guidelines 3 Calibration Resources 4 Required Software 4 Recommended...

Page 3: ...s the measured calibration results against the defined As Found Limits The result of the evaluation is expressed as a Pass Fail condition in the calibration certificate using an established evaluation...

Page 4: ...ons 41 points Adjust only 5 minutes 12 connections 6 points Note Estimated test times assume the user is conducting a manual calibration For most procedures automating the calibration significantly re...

Page 5: ...er adjustment ensure that the Test Conditions have been met before returning the DUT to NI Allow the calibrator and DMM to settle fully before taking any measurements NI recommends a settling time of...

Page 6: ...pment NI recommends for each test performed as part of this calibration procedure NI Calibration Executive Users Refer to the Calibration Executive Help to find an updated list of test equipment for t...

Page 7: ...fications DC Gain AC Gain BNC f to Banana Plug Adaptor Pomona Model 1269 Verifications DC Gain AC Gain Power Supply NI PS 16 Verifications DC Gain AC Gain 24 V DC 5 A BNC f to BNC f cable Verification...

Page 8: ...rovides additional correction for errors that result from measurement such as temperature drift Refer to the DMM user documentation for more information on conducting ACAL Account for additional uncer...

Page 9: ...imit Lower Limit Upper Limit 1000 V 5 V 1100 V 10 V 502 mV 502 mV 251 mV 251 mV 750 V 3 75 V 1100 V 10 V 377 mV 377 mV 189 mV 189 mV 500 V 2 5 V 1100 V 10 V 252 mV 252 mV 126 mV 126 mV 200 V 1 V 220 V...

Page 10: ...ck Power off all high voltage signals before making any connections Figure 1 DC Gain Connection 1 Calibrator 7 BNC 2090A 2 Low thermal EMF cable with banana plugs 8 PXIe 6356 3 RM 26999 9 SH68F 68F EP...

Page 11: ...connections 4 Connect the calibrator output to the appropriate channel on the DUT Wait 2 minutes for thermal EMF to stabilize Repeat 13 times once for each set of voltage input values 5 Configure the...

Page 12: ...ement with the DMM 10 Scale the DMM measurement with the coefficients stored on the DUT EEPROM a RM 26999 Initialize Voltage Channel vi b RM 26999 lib _getChannelsFromTask vi c RM 26999 lib _filterPhy...

Page 13: ...V 113 mVRMS 113 mVRMS 55 mVRMS 55 mVRMS 5 VRMS 0 025 VRMS 22 V 0 1 V 80 5 mVRMS 80 5 mVRMS 40 mVRMS 40 mVRMS 5 kHz 1000 VRMS 5 VRMS 1100 V 10 V 9 0 VRMS 9 0 VRMS 4 5 VRMS 4 5 VRMS 500 VRMS 2 5 VRMS 11...

Page 14: ...50 VRMS 0 25 VRMS 22 V 1 V 688 mVRMS 688 mVRMS 340 mVRMS 340 mVRMS 5 VRMS 0 025 VRMS 22 V 0 1 V 80 5mVRMS 80 5 mVRMS 40 mVRMS 40 mVRMS 1 MHz 220 VRMS 1 1 VRMS 220 V 1 V 30 8 VRMS 30 8 VRMS 15 4 VRMS 1...

Page 15: ...hock Power off all high voltage signals before making any connections Figure 2 AC Gain Connection 1 Calibrator 7 NI BNC 2090A 2 Calibrator amp cable 8 BNC cable 3 Calibrator amp 9 BNC to banana plug a...

Page 16: ...setting 4 Set the calibrator and the DMM to the appropriate frequency defined in Table 2 Repeat 4 times once for each set of voltage input values 5 Configure the DMM to measure the appropriate low vo...

Page 17: ...with the DMM 10 Scale the DMM measurement with the coefficients stored on the DUT EEPROM a RM 26999 Initialize Voltage Channel vi b RM 26999 lib _getChannelsFromTask vi c RM 26999 lib _filterPhysical...

Page 18: ...date and the temperature in the DUT EEPROM If the DUT passes the verification procedures within the As Left test limits an adjustment is not required Note NI recommends an adjustment to optimize the m...

Page 19: ...dure 18 ni com calibration Figure 3 DC Gain Connection 1 Calibrator 6 BNC cable 2 Low thermal EMF cable with banana plugs 7 BNC to banana plug adaptor 3 NI RM 26999 8 DMM 4 NI SH68F 68F EPM cable 9 Po...

Page 20: ...ibration Stimulus vi a Use the high voltage values from Table 3 as input for the VI Repeat 2 times once for each polarity Repeat 4 times once for each channel 4 Power off the calibrator output Caution...

Page 21: ...ain Calibration vi a Use the high voltage value as the reference voltage signal as defined in Table 3 b Use the low voltage value as the measured voltage signal as defined in Table 3 Repeat 4 times on...

Page 22: ...2 times once for each signal path Repeat 1 time once for each frequency Table 5 AC Voltage Test Point Frequency High Voltage Value Calibrator Range Low Voltage Value DMM Range 15 kHz 10 VRMS 22 VRMS...

Page 23: ...vi with the commit action to store all scaling coefficients for both DC and AC on the DUT EEPROM Perform Reverification Perform all tests in the Verification section after completing Adjustment This v...

Page 24: ...companies For patents covering National Instruments products technology refer to the appropriate location Help Patents in your software the patents txt file on your media or the National Instruments...

Reviews: