background image

V — Compare high or low, not midband; store results from capture functionality if
configured.

D — Drive data from source functionality if configured.

M — Compare midband, not high or low.

Note

 

Termination mode settings affect the termination applied to all non-driving

pin states. Non-drive states include L, H, M, V, X and potentially -. Refer to the 

Programmable input termination mode

 specification for more information.

Edge Timing

Edge Types

Drive edges

4; drive on, drive data, drive return, drive off

Compare edge

1; strobe

Number of time sets

4

31

Edge Generation Timing

Edge placement range

Minimum

Start of vector period (0 ns)

Maximum

5 vector periods or 40 µs, whichever is smaller

Minimum required edge separation

Between any driven data change

5 ns

Between any Drive On and Drive Off
edges

5 ns

Between Compare Strobes

5 ns

Edge placement resolution

39.0625 ps

Edge placement accuracy

5

Drive

±500 ps, warranted

Compare

±500 ps, warranted

Overall timing accuracy

5

±1.5 ns, warranted

TDR deskew adjustment resolution

39.0625 ps

4

31 time sets can be configured. One additional time set, represented by a -, repeats the previous
time set.

5

For specifications across multiple instruments in a Semiconductor Test System, refer to the

Semiconductor Test System Specifications

.

PXIe-6570 Specifications

  | 

© National Instruments

  | 

5

Summary of Contents for PXIe-6570

Page 1: ...ics are unwarranted values that are relevant to the use of the product and convey the expected performance of the product Specifications in this document are Characteristics unless otherwise noted Spe...

Page 2: ...Voltage 7 PPMU Measure Voltage 7 PPMU Force Current 7 PPMU Measure Current 8 PPMU Programmable Aperture Time 9 Pattern Control 10 Opcodes 10 Pipeline Latencies 11 Source and Capture 12 Frequency Measu...

Page 3: ...e fly Clock Generation Master clock source PXIe_CLK1003 Clock domains One independent clock domains on a single instrument not supported 1 The system channel count is the maximum number of channels av...

Page 4: ...te State Non Return NR Pin State Vector Cycle Drive Data Drive Return Drive Data Drive On Drive On Drive Return Drive Data Drive Return Drive Off Drive On Drive Data Drive On Drive Return Drive Data D...

Page 5: ...Timing Edge placement range Minimum Start of vector period 0 ns Maximum 5 vector periods or 40 s whichever is smaller Minimum required edge separation Between any driven data change 5 ns Between any...

Page 6: ...tput impedance 50 Rise fall time 20 to 80 1 2 ns up to 5 V Comparator Signal type Single ended referenced to the DGS pin when connected Otherwise referenced to GND Programmable levels VOH VOL Voltage...

Page 7: ...Accuracy 15 mV 1 M load warranted PPMU Measure Voltage Signal type Single ended referenced to the DGS pin when connected Otherwise referenced to GND Voltage levels Range 2 V to 6 V Resolution 228 V A...

Page 8: ...PMU voltage clamps Range 2 V to 6 V Resolution 122 V Accuracy 100 mV typical PPMU Measure Current Table 2 PPMU Measure Current Accuracy Range Resolution Accuracy 2 A 460 pA 1 of range for Zone 1 of Fi...

Page 9: ...Current Percent of Current Range Measured Voltage V 25 75 50 25 0 50 75 1 2 3 1 0 4 5 100 2 100 6 Zone 2 Zone 1 1 25 1 75 PPMU Programmable Aperture Time Aperture time Minimum 4 s Maximum 65 ms Resol...

Page 10: ...0 100m 190 150 140 10m Pattern Control Opcodes Refer to the following table for supported opcodes Using matched and failed opcode parameters with multiple PXIe 6570 instruments requires the PXIe 6674T...

Page 11: ...ear_seqflag write_reg Signal set_signal pulse_signal clear_signal Digital Source and Capture capture_start capture capture_stop source_start source Pipeline Latencies Minimum delay between source_star...

Page 12: ...g equation to calculate the frequency counter error ppm 1 20 1 000 000 where MeasurementTime is the time in seconds over which the frequency counter measurement is configured to run UnknownClockPeriod...

Page 13: ...with a 1 ms measurement time using the PXIe Clk100 provided by the PXIe 1085 chassis as the timebase Solution 25 1 25 20 1 100 1 000 000 45 Example 2 Calculating Error when Overriding with the PXIe 66...

Page 14: ...r rails The maximum current drawn from each of these rails can vary depending on the PXIe 6570 mode of operation The total power consumption does not exceed the input power specification Input power 6...

Page 15: ...60068 2 64 Nonoperating test profile exceeds the requirements of MIL PRF 28800F Class 3 Safety Voltages Connect only voltages that are within these limits Supported measurement range 2 V to 6 V Cauti...

Page 16: ...spection analysis purposes Note For EMC declarations certifications and additional information refer to the Online Product Certification section CE Compliance This product meets the essential requirem...

Page 17: ...cycle all NI products must be disposed of according to local laws and regulations For more information about how to recycle NI products in your region visit ni com environment weee RoHS National Inst...

Page 18: ...legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relevant...

Reviews: