![background image](http://html1.mh-extra.com/html/national-instruments/pxie-5774/pxie-5774_getting-started-manual_3622830010.webp)
Signal
Type
Direction
DIO <0..7>
Single-ended
Bidirectional
5.0 V
DC
Output
GND
Ground
—
Notice
The maximum input signal levels are valid only when the module is
powered on. To avoid permanent damage to the PXIe-5774, do not apply a signal to
the device when the module is powered down.
Notice
Connections that exceed any of the maximum ratings of any connector on
the PXIe-5774 can damage the device and the system. NI is not liable for any
damage resulting from such connections.
Configuring the PXIe-5774 in MAX
Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX
informs other programs about which NI hardware products are in the system and how they are
configured. MAX is automatically installed with FlexRIO.
1.
Launch MAX.
2.
In the configuration tree, expand
Devices and Interfaces
to see the list of installed NI
hardware.
Installed modules appear under the name of their associated chassis.
3.
Expand your
Chassis
tree item.
MAX lists all modules installed in the chassis. Your default names may vary.
Note
If you do not see your module listed, press <F5> to refresh the list of
installed modules. If the module is still not listed, power off the system, ensure
the module is correctly installed, and restart.
4.
Record the identifier MAX assigns to the hardware. Use this identifier when
programming the PXIe-5774.
5.
Self-test the hardware by selecting the item in the configuration tree and clicking
Self-
Test
in the MAX toolbar.
The MAX self-test performs a basic verification of hardware resources.
FlexRIO Examples
FlexRIO includes several example applications for LabVIEW. These examples serve as
interactive tools, programming models, and as building blocks in your own applications.
10
|
ni.com
|
PXIe-5774 Getting Started Guide