All parameters are set to identical values for each module.
•
Modules are synchronized without using an external Sample Clock.
Note
Although you can use NI-TClk to synchronize non-identical SMC-based
modules, these specifications apply only to synchronizing identical modules.
Skew
16
AC-coupled
130 ps, measured
DC-coupled
140 ps, measured
Skew after manual adjustment
≤10 ps, measured
Sample Clock delay/adjustment
1.5 ps
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety
standards for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA C22.2 No. 61010-1
Note
For UL and other safety certifications, refer to the product label or the
Online
Product Certification
section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment
for measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
EN 55022 (CISPR 22): Class A emissions
•
EN 55024 (CISPR 24): Immunity
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
AS/NZS CISPR 22: Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note
In the United States (per FCC 47 CFR), Class A equipment is intended for
use in commercial, light-industrial, and heavy-industrial locations. In Europe,
16
Caused by clock and analog delay differences. No manual adjustment performed. Tested with a
PXIe-1085 chassis with a 24 GB backplane with a maximum slot to slot skew of 100 ps. Measured
at 23 °C.
PXIe-5763 Specifications
|
© National Instruments
|
19