background image

Programming the NI 5622

You can acquire data interactively using the NI-SCOPE SFP, or you can use the NI-SCOPE
instrument driver to program your device in the supported ADE of your choice.

Table 2.

 NI 5622 Programming Options

Application

Programming

Interface (API)

Location

Description

 

 

 

NI-SCOPE SFP

Available from the start menu at

Start

»

All Programs

»

National

Instruments

»

NI-SCOPE

»

NI-

SCOPE Soft Front Panel

.

The NI-SCOPE SFP acquires,
controls, analyzes, and presents
data, similar to stand-alone
oscilloscopes. The NI-SCOPE SFP
operates on the PC, so you can view
and control waveforms directly
from your computer. You can also
run multiple sessions of the
NI-SCOPE SFP simultaneously.

NI-SCOPE
Instrument Driver

LabVIEW—Available on the
LabVIEW Functions palette at

Measurement I/O

»

NI-SCOPE

.

The NI-SCOPE API configures and
operates the device hardware and
provides customizable acquisition,
control, analysis, and measurement
options using LabVIEW VIs or
LabWindows/CVI functions.

C or LabWindows/CVI—
Available at 

Program Files

»

IVI

Foundation

»

IVI

»

Drivers

»

niScope

.

Microsoft Visual C/C++—Use
examples located in the

<NIDocDir>\NI–SCOPE
\examples

 directory, where

<NIDocDir>

 is one of the

following directories:

Windows 8/7/Vista—

Users

\Public\Documents
\National
Instruments

Windows XP—

Documents

and Settings\All
Users\Shared
Documents\National
Instruments

You can modify an NI-SCOPE C
example to create an application
with Microsoft Visual C/C++. Copy
an NI-SCOPE example to copy
required project settings for include
paths and library files. Alternatively,
refer to the 

Creating an Application

with Microsoft Visual C and C++

topic of the 

NI High-Speed

Digitizers Help

 to manually add all

required include and library files to
your project.

8

  | 

ni.com

  | 

NI PXIe-5622 Getting Started Guide

Summary of Contents for PXIe-5622

Page 1: ...nt 3 Preparing the Environment 3 Installing the Software 4 Installing the NI 5622 4 Hardware Front Panel Connectors 6 Configuring the Hardware in MAX 7 Programming the NI 5622 8 Troubleshooting 9 Why...

Page 2: ...cables must be no longer than 30 m 100 ft Verifying the System Requirements To use the NI 5622 your system must meet certain requirements For more information about minimum system requirements recomm...

Page 3: ...n are required for installation and operation of the NI 5622 Preparing the Environment Ensure that the environment in which you are using the NI 5622 meets the following specifications Operating Envir...

Page 4: ...Accept the prompts to complete the installation 4 When the installer completes select Restart in the dialog box that prompts you to restart shut down or restart later Installing the NI 5622 Caution T...

Page 5: ...til it is fully inserted Figure 3 Module Installation NI PXIe 1075 2 3 1 1 Chassis 2 Hardware Module 3 Ejector Handle in Down Unlatched Position 9 Latch the module in place by pulling up on the ejecto...

Page 6: ...ACTIVE PFI 1 IF IN CLK IN CLK OUT TTL 5 V MAX Table 1 NI 5622 Front Panel Signal Connections Connector Description Function IF IN SMA connector Analog input connection digitizes data and triggers acq...

Page 7: ...ices If the device is still not listed power off the system ensure the device is correctly installed and restart 4 Record the device identifier MAX assigns to the hardware Use this identifier when pro...

Page 8: ...COPE The NI SCOPE API configures and operates the device hardware and provides customizable acquisition control analysis and measurement options using LabVIEW VIs or LabWindows CVI functions C or LabW...

Page 9: ...rent chassis slot 5 Power on the chassis 6 Verify that the device appears in MAX 7 Reset the device in MAX and perform a self test Related Information Installing the NI 5622 on page 4 What Should I Do...

Page 10: ...dule in a different slot 5 Power on the chassis 6 Perform the self test again Thermal Shutdown Error If you receive an over temperature or thermal shutdown error and your device shuts down complete th...

Page 11: ...Help Worldwide Support and Services The National Instruments website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and applicati...

Page 12: ...mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products technology refer to the appropriate location Help Patents in your softwa...

Reviews: