background image

Ultra

compact sensor for smallest quantities, with

a transmitter remote version with up to 4 I/Os

Application

• Measuring principle operates independently of physical fluid

properties such as viscosity or density

• Measuring accurately smallest quantities of liquids and

gases

Device properties

• Nominal diameter: DN 1 to 6 (¹⁄₂₄ to ¹⁄₄")

• Process pressure up to 400 bar (5

 

800 psi)

• Medium temperature up to +205 °C (+401 °F)

• Remote version with up to 4 I/Os

• Backlit display with touch control and WLAN access

• Standard cable between sensor and transmitter

Your benefits

• Reduced Installation cost – compact single-tube design

• Fewer process measuring points – multivariable

measurement (flow, density, temperature)

• Space

saving installation – no inlet/outlet run needs

• Full access to process and diagnostic information –

numerous, freely combinable I/Os and fieldbuses

• Reduced complexity and variety – freely configurable I/O

functionality

• Integrated verification – Heartbeat Technology

Products

Solutions

Services

Technical Information

Proline Cubemass C 500

Coriolis flowmeter

TI01281D/06/EN/03.18

71402594

2018-06-01

Заказывайте

 

на

 

сайте

: www.tinnova.ru/endress_hauser/ || 

Эл

почта

[email protected]

Summary of Contents for Proline Cubemass C 500

Page 1: ... access Standard cable between sensor and transmitter Your benefits Reduced Installation cost compact single tube design Fewer process measuring points multivariable measurement flow density temperature Space saving installation no inlet outlet run needs Full access to process and diagnostic information numerous freely combinable I Os and fieldbuses Reduced complexity and variety freely configurab...

Page 2: ...mitter housing 55 Special mounting instructions 56 Environment 59 Ambient temperature range 59 Storage temperature 59 Climate class 59 Degree of protection 60 Vibration resistance 60 Shock resistance 60 Shock resistance 60 Interior cleaning 60 Electromagnetic compatibility EMC 60 Process 60 Medium temperature range 60 Density 61 Pressure temperature ratings 61 Sensor housing 63 Rupture disk 64 Flo...

Page 3: ...A server 103 Accessories 104 Device specific accessories 104 Communication specific accessories 105 Service specific accessories 106 System components 106 Supplementary documentation 107 Standard documentation 107 Device dependent additional documentation 107 Registered trademarks 108 Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 4: ...ns supply Outer ground terminal Connects the device to the plant grounding system Communication symbols Symbol Meaning Wireless Local Area Network WLAN Communication via a wireless local network LED Light emitting diode is off LED Light emitting diode is on LED Light emitting diode is flashing Symbols for certain types of information Symbol Meaning Permitted Procedures processes or actions that ar...

Page 5: ...nt rotational velocity ω the sensor uses oscillation In the sensor an oscillation is produced in the measuring tube The Coriolis forces produced at the measuring tube loop cause a phase shift in the tube oscillations see illustration If there is zero flow i e when the fluid stands still the oscillation measured at points A and B has the same phase no phase difference 1 Mass flow causes deceleratio...

Page 6: ...ncy is thus a function of medium density The microprocessor utilizes this relationship to obtain a density signal Volume measurement Together with the measured mass flow this is used to calculate the volume flow Temperature measurement The temperature of the measuring tube is determined in order to calculate the compensation factor due to temperature effects This signal corresponds to the process ...

Page 7: ...ection housing Signal transmission digital Order code for Integrated ISEM electronics option A Sensor Electronics and ISEM intelligent sensor electronics module in the transmitter housing Signal transmission analog Order code for Integrated ISEM electronics option B Transmitter Connecting cable can be ordered in various lengths 104 Length Zone 2 Class I Division 2 max 300 m 1000 ft Zone 1 Class I ...

Page 8: ...ic version stainless steel 1 4301 304 Optional order code for Sensor feature option CC Hygienic version for maximum corrosion resistance stainless steel 1 4404 316L This device version is only available in conjunction with the Proline 500 digital transmitter Order code for Sensor connection housing option L Cast stainless 1 4409 CF3M similar to 316L Sensor Cubemass C A0029825 The ultra compact sin...

Page 9: ...tive measures on the operator s side These functions can be configured by the user and guarantee greater in operation safety if used correctly An overview of the most important functions is provided in the following section Function interface Factory setting Recommendation Write protection via hardware write protection switch 10 Not enabled On an individual basis following risk assessment Access c...

Page 10: ...y The WLAN authentication of the network key complies with the IEEE 802 11 standard When the device is delivered the network key is pre defined depending on the device It can be changed via the WLAN settings submenu in the WLAN passphrase parameter Infrastructure mode A connection between the device and WLAN access point is protected by means of an SSID and passphrase on the system side Please con...

Page 11: ...terface CDI RJ45 The device can be connected to a network via the service interface CDI RJ45 Device specific functions guarantee the secure operation of the device in a network The use of relevant industrial standards and guidelines that have been defined by national and international safety committees such as IEC ISA62443 or the IEEE is recommended This includes organizational security measures s...

Page 12: ... the sound velocity of the gas used and can be calculated with the formula below max G minimum max F ρG x ρG cG π 2 di 2 3600 max G Maximum full scale value for gas kg h max F Maximum full scale value for liquid kg h max G max F max G can never be greater than max F ρG Gas density in kg m at operating conditions x Constant dependent on nominal diameter cG Sound velocity gas m s di Measuring tube i...

Page 13: ...measured values to calculate the corrected volume flow HART protocol The measured values are written from the automation system to the measuring device via the HART protocol The pressure transmitter must support the following protocol specific functions HART protocol Burst mode Current input The measured values are written from the automation system to the measuring device via the current input 13...

Page 14: ...put signal level Low signal DC 3 to 5 V High signal DC 12 to 30 V Assignable functions Off Reset the individual totalizers separately Reset all totalizers Flow override Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 15: ... 20 mA Ex i C C C User configurable input output 1 D D D D D D D Pulse frequency switch output E E E E E E E Double pulse output 2 F F Pulse frequency switch output Ex i G G G Relay output H H H H H H H Current input 0 4 to 20 mA I I I I I I I Status input J J J J J J J Order code for Output input 3 022 Output input 4 023 3 Not assigned A A A A A A A A A A Current output 0 4 to 20 mA B B B B B Cur...

Page 16: ...y Exciter current 0 The range of options increases if the measuring device has one or more application packages PROFIBUS PA PROFIBUS PA In accordance with EN 50170 Volume 2 IEC 61158 2 MBP galvanically isolated Data transmission 31 25 kbit s Current consumption 10 mA 16 mA Permitted supply voltage 9 to 32 V Bus connection With integrated reverse polarity protection PROFIBUS DP Signal encoding NRZ ...

Page 17: ...µA Damping Adjustable 0 07 to 999 s Assignable measured variables Mass flow Volume flow Corrected volume flow Density Reference density Temperature Electronic temperature Oscillation frequency 0 Oscillation damping 0 Signal asymmetry Exciter current 0 The range of options increases if the measuring device has one or more application packages Pulse frequency switch output Function Can be set to pul...

Page 18: ...ture Electronic temperature Oscillation frequency 0 Oscillation damping 0 Signal asymmetry Exciter current 0 The range of options increases if the measuring device has one or more application packages Switch output Maximum input values DC 30 V 250 mA passive Open circuit voltage DC 28 8 V active Switching behavior Binary conductive or non conductive Switching delay Adjustable 0 to 100 s Number of ...

Page 19: ...ctory setting NC normally closed Maximum switching capacity passive DC 30 V 0 1 A AC 30 V 0 5 A Assignable functions Off On Diagnostic behavior Limit value Mass flow Volume flow Corrected volume flow Density Reference density Temperature Totalizer 1 3 Flow direction monitoring Status Partially filled pipe detection Low flow The range of options increases if the measuring device has one or more app...

Page 20: ...herNet IP Device diagnostics Device condition can be read out in Input Assembly PROFINET Device diagnostics According to Application Layer protocol for decentralized periphery Version 2 3 FOUNDATION Fieldbus Status and alarm messages Diagnostics in accordance with FF 891 Error current FDE Fault Disconnection Electronic 0 mA Modbus RS485 Failure mode Choose from NaN value instead of current value L...

Page 21: ... Current status Open Closed Local display Plain text display With information on cause and remedial measures Backlight Red backlighting indicates a device error Status signal as per NAMUR recommendation NE 107 Interface protocol Via digital communication HART protocol FOUNDATION Fieldbus PROFIBUS PA PROFIBUS DP Modbus RS485 EtherNet IP PROFINET Via service interface CDI RJ45 service interface WLAN...

Page 22: ...M 250 VAC Option MA Modbus RS485 UN 30 VDC UM 250 VAC Option SA FOUNDATION Fieldbus UN 30 VDC UM 250 VAC Option NA EtherNet IP UN 30 VDC UM 250 VAC Option RA PROFINET UN 30 VDC UM 250 VAC Order code for Output input 2 Output input 3 Output input 4 Output type Safety related values Output input 2 Output input 3 Output input 4 1 24 25 22 23 20 21 Option B Current output 4 to 20 mA UN 30 VDC UM 250 V...

Page 23: ...tput type Intrinsically safe values or NIFW values Output input 2 Output input 3 Output input 4 1 24 25 22 23 20 21 Option C Current output 4 to 20 mA Ex i Ui 30 V li 100 mA Pi 1 25 W Li 0 Ci 0 Option G Pulse frequency switch output Ex i Ui 30 V li 100 mA Pi 1 25 W Li 0 Ci 0 1 The order code Output input 4 is only available for the Proline 500 digital transmitter Low flow cut off The switch points...

Page 24: ... the Promass 500 GSD file Earlier models Promass 80 PROFIBUS PA ID No 1528 hex Extended GSD file EH3x1528 gsd Standard GSD file EH3_1528 gsd Promass 83 PROFIBUS PA ID No 152A hex Extended GSD file EH3x152A gsd Standard GSD file EH3_152A gsd Description of the function scope of compatibility Operating Instructions 107 System integration Information regarding system integration Operating Instruction...

Page 25: ...are FieldCare Add on Profile Level 3 for Rockwell Automation control systems Web browser Electronic Data Sheet EDS integrated in the measuring device Configuration of the EtherNet interface Speed 10 MBit 100 MBit auto factory setting Duplex half duplex full duplex auto factory setting Configuration of the device address DIP switches on the electronics module for IP addressing last octet DHCP Manuf...

Page 26: ... name assignment last part DCP protocol Process Device Manager PDM Integrated Web server Supported functions Identification Maintenance Simple device identification via Control system Nameplate Measured value status The process variables are communicated with a measured value status Blinking feature via the onsite display for simple device identification and assignment Device operation via operati...

Page 27: ...ystem integration Operating Instructions 107 Cyclic data transmission Description of the modules Execution times Methods Modbus RS485 Protocol Modbus Applications Protocol Specification V1 1 Response times Direct data access typically 25 to 50 ms Auto scan buffer data range typically 3 to 5 ms Device type Slave Slave address range 1 to 247 Broadcast address range 0 Function codes 03 Read holding r...

Page 28: ...ation Information on system integration Operating Instructions 107 Modbus RS485 information Function codes Register information Response time Modbus data map Power supply Terminal assignment Transmitter supply voltage input outputs HART Supply voltage Input output 1 Input output 2 Input output 3 Input output 4 1 2 26 27 24 25 22 23 20 21 The terminal assignment depends on the specific device versi...

Page 29: ...epends on the specific device version ordered 15 Transmitter and sensor connection housing connecting cable The sensor and transmitter which are mounted in separate locations are interconnected by a connecting cable The cable is connected via the sensor connection housing and the transmitter housing Terminal assignment and connection of the connecting cable Proline 500 digital 33 Proline 500 34 De...

Page 30: ...t IP Order code for Cable entry connection 34 Electrical connection 2 3 L N P U Connector M12 1 R 1 2 S 1 2 T 1 2 V 1 2 Connector M12 1 Connector M12 1 1 Cannot be combined with an external WLAN antenna order code for Enclosed accessories option P8 of an RJ45 M12 adapter for the service interface order code for Accessories mounted option NB or of the remote display and operating module DKX001 2 Su...

Page 31: ... 04 Phoenix part no 1543223 SACC M12MSD 4Q When using the device in a hazardous location use a suitably certified plug Service interface Order code for Accessories mounted option NB Adapter RJ45 M12 service interface 3 2 4 1 A0032047 Pin Assignment 1 Tx 2 Rx 3 Tx 4 Rx Coding Plug socket D Socket Recommended plug Binder series 763 part no 99 3729 810 04 Phoenix part no 1543223 SACC M12MSD 4Q When u...

Page 32: ...consumption Transmitter Max 10 W active power Current consumption Transmitter Max 400 mA 24 V Max 200 mA 110 V 50 60 Hz 230 V 50 60 Hz Power supply failure Depending on the device version the configuration is retained in the device memoryor in the pluggable data memory HistoROM DAT Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 33: ...rder code for Housing Connection to sensor connection housing via Connection to transmitter housing via Option A aluminum coated Terminals Terminals Option B stainless Terminals Terminals Option C ultra compact hygienic stainless Device plug Terminals Option L cast stainless Terminals Terminals Pin assignment device plug Device plugs are only available for device version order code for Housing Opt...

Page 34: ...ection for signal transmission input output 3 Terminal connection for signal transmission input output 4 Terminal connection for connecting cable between sensor and transmitter 5 Terminal connection for signal transmission input output or terminal for network connection DHCP client via service interface CDI RJ45 optional terminal connection for external WLAN antenna 6 Protective ground PE An adapt...

Page 35: ...M12 connector without opening the device Network connection DHCP client via service interface CDI RJ45 93 Connecting in a ring topology Device versions with EtherNet IP and PROFINET communication protocols can be integrated into a ring topology The device is integrated via the terminal connection for signal transmission output 1 and the connection to the service interface CDI RJ45 Integrating the ...

Page 36: ... active 1 Automation system with current input e g PLC 2 Cable shield the cable shield must be grounded at both ends to comply with EMC requirements observe cable specifications 44 3 Connection for HART operating devices 87 4 Resistor for HART communication 250 Ω observe maximum load 16 5 Analog display unit observe maximum load 16 6 Transmitter 2 3 4 20 mA 4 1 5 A0028762 3 Connection example for ...

Page 37: ...ications 4 Analog display unit observe maximum load 5 Pressure transmitter e g Cerabar M Cerabar S see requirements 6 Transmitter PROFIBUS PA 2 1 3 4 7 8 6 6 6 6 5 6 6 5 A0028768 5 Connection example for PROFIBUS PA 1 Control system e g PLC 2 PROFIBUS PA segment coupler 3 Cable shield the cable shield must be grounded at both ends to comply with EMC requirements observe cable specifications 4 T bo...

Page 38: ... EMC requirements observe cable specifications 3 Distribution box 4 Transmitter If baud rates 1 5 MBaud an EMC cable entry must be used and the cable shield must continue as far as the terminal wherever possible EtherNet IP 1 2 4 3 5 5 A0028767 7 Connection example for EtherNet IP 1 Control system e g PLC 2 Ethernet switch 3 Observe cable specifications 4 Device plug 5 Transmitter Заказывайте на с...

Page 39: ...OFINET 1 2 4 5 5 3 A0016805 8 Connection example for PROFINET 1 Control system e g PLC 2 Ethernet switch 3 Observe cable specifications 4 Device plug 5 Transmitter PROFINET MRP Media Redundancy Protocol 1 2 4 5 5 3 A0027544 1 Control system e g PLC 2 Ethernet switch 3 Observe cable specifications 44 4 Connecting cable between the two transmitters 5 Transmitter Заказывайте на сайте www tinnova ru e...

Page 40: ...bserve cable specifications 4 T box 5 Measuring device 6 Local grounding 7 Bus terminator 8 Potential matching line Modbus RS485 2 1 A B 3 4 4 A B A B A0028765 10 Connection example for Modbus RS485 non hazardous area and Zone 2 Class I Division 2 1 Control system e g PLC 2 Cable shield the cable shield must be grounded at both ends to comply with EMC requirements observe cable specifications 3 Di...

Page 41: ...ction example for 4 20 mA current output passive 1 Automation system with current input e g PLC 2 Active barrier for power supply e g RN221N 3 Analog display unit observe maximum load 4 Transmitter Pulse frequency output 1 2 3 12345 A0028761 13 Connection example for pulse frequency output passive 1 Automation system with pulse frequency input e g PLC 2 Power supply 3 Transmitter Observe input val...

Page 42: ...ulse output active 1 Automation system with double pulse input e g PLC 2 Transmitter Observe input values 19 3 Double pulse output 4 Double pulse output slave phase shifted 1 3 2 4 5 A0029279 16 Connection example for double pulse output passive 1 Automation system with double pulse input e g PLC 2 Power supply 3 Transmitter Observe input values 19 4 Double pulse output 5 Double pulse output slave...

Page 43: ... for 4 to 20 mA current input 1 Power supply 2 Terminal box 3 External measuring device for reading in pressure or temperature for instance 4 Transmitter Status input 1 2 3 A0028764 19 Connection example for status input 1 Automation system with status output e g PLC 2 Power supply 3 Transmitter Potential equalization Requirements No special measures for potential equalization are required Заказыв...

Page 44: ...inimum and maximum temperatures to be expected Power supply cable Standard installation cable is sufficient Protective ground cable Cable 2 08 mm2 14 AWG The grounding impedance must be less than 1 Ω Signal cable Current output 4 to 20 mA HART A shielded cable is recommended Observe grounding concept of the plant PROFIBUS PA Twisted shielded two wire cable Cable type A is recommended For further i...

Page 45: ...ROFINET networks see PROFINET Cabling and Interconnection Technology Guideline for PROFINET FOUNDATION Fieldbus Twisted shielded two wire cable For further information on planning and installing FOUNDATION Fieldbus networks see Operating Instructions for FOUNDATION Fieldbus Overview BA00013S FOUNDATION Fieldbus Guideline IEC 61158 2 MBP Modbus RS485 The EIA TIA 485 standard specifies two types of ...

Page 46: ...ass I Division 2 sensor installed in the hazardous area Zone 2 Class I Division 2 B Standard cable to 500 digital transmitter 47 Transmitter installed in the hazardous area Zone 2 Class I Division 2 sensor installed in the hazardous area Zone 1 Class I Division 1 C Signal cable to 500 transmitter 49 Transmitter and sensor installed in the hazardous area Zone 2 Class I Division 2 oder Zone 1 Class ...

Page 47: ...ixed 20 m 65 ft variable up to maximum 50 m 165 ft 1 UV radiation can impair the cable outer sheath Protect the cable from direct sunshine where possible B Connecting cable between sensor and transmitter Proline 500 digital Standard cable A standard cable with the following specifications can be used as the connecting cable Design 4 6 8 cores 2 3 4 pairs uninsulated stranded CU wires pair stranded...

Page 48: ... cable for Zone 1 Class I Division 1 Standard cable 2 2 0 5 mm2 AWG 20 PVC cable 1 with common shield 2 pairs pair stranded Flame resistance According to DIN EN 60332 1 2 Oil resistance According to DIN EN 60811 2 1 Shielding Tin plated copper braid optical cover 85 Operating temperature When mounted in a fixed position 50 to 105 C 58 to 221 F when cable can move freely 25 to 105 C 13 to 221 F Ava...

Page 49: ...rror limits based on ISO 11631 Water with 15 to 45 C 59 to 113 F at2 to 6 bar 29 to 87 psi Specifications as per calibration protocol Accuracy based on accredited calibration rigs that are traced to ISO 17025 To obtain measured errors use the Applicator sizing tool 106 Maximum measured error o r of reading 1 g cm 1 kg l T medium temperature Base accuracy Design fundamentals 52 Mass flow and volume...

Page 50: ... 0 368 0 074 Accuracy of outputs The outputs have the following base accuracy specifications Current output Accuracy 5 µA Pulse frequency output o r of reading Accuracy Max 50 ppm o r over the entire ambient temperature range Repeatability o r of reading 1 g cm3 1 kg l T medium temperature Base repeatability Design fundamentals 52 Mass flow and volume flow liquids 0 05 o r Mass flow gases 0 25 o r...

Page 51: ...0025 g cm3 F Field density calibration is possible Wide range density specification special density calibration If the process temperature is outside the valid range 49 the measured error is 0 00005 g cm3 C 0 000025 g cm3 F kg m 3 C 40 0 80 40 80 120 160 200 240 280 320 360 400 F 50 0 50 100 150 200 0 2 4 6 8 10 2 1 A0016617 1 Field density calibration for example at 20 C 68 F 2 Special density ca...

Page 52: ...021334 Calculation of the maximum repeatability as a function of the flow rate Flow rate Maximum repeatability in o r ZeroPoint BaseRepeat 100 A0021335 A0021340 ZeroPoint BaseRepeat 100 A0021336 ZeroPoint MeasValue 100 A0021337 Example for maximum measured error 0 10 20 30 40 50 60 0 0 5 1 0 1 5 2 0 2 5 70 80 90 100 Q E A0030288 E Maximum measured error in o r example Q Flow rate in of maximum ful...

Page 53: ...ameter prevent the sensor running empty while measurement is in progress 1 2 3 4 5 A0028773 20 Installation in a down pipe e g for batching applications 1 Supply tank 2 Sensor 3 Orifice plate pipe restriction 4 Valve 5 Batching tank DN Ø orifice plate pipe restriction mm in mm in 1 0 8 0 03 2 1 5 0 06 4 3 0 0 12 6 5 0 0 20 Orientation The direction of the arrow on the sensor nameplate helps you to...

Page 54: ...ient temperature for the transmitter this orientation is recommended 2 Applications with high process temperatures may increase the ambient temperature To maintain the maximum ambient temperature for the transmitter this orientation is recommended Inlet and outlet runs No special precautions need to be taken for fittings which create turbulence such as valves elbows or T pieces as long as no cavit...

Page 55: ... A0029051 21 Engineering unit mm in Wall mounting 149 5 85 211 8 31 5 8 0 23 17 0 67 L 5 8 0 23 A0029054 22 Engineering unit mm in L Depends on order code for Transmitter housing Order code for Transmitter housing Option A aluminum coated L 14 mm 0 55 in Option D polycarbonate L 13 mm 0 51 in Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 56: ...on a secure fixed post Only mount the transmitter on a secure fixed post on a stable surface 1 4 x SW 13 0 79 to 2 75 20 70 A0029057 23 Engineering unit mm in Wall mounting 20 0 79 100 3 94 100 3 94 8 6 0 39 18 0 71 10 0 39 A0029068 24 Engineering unit mm in Special mounting instructions Rupture disk Information that is relevant to the process 64 Заказывайте на сайте www tinnova ru endress_hauser ...

Page 57: ...failure of the rupture disk a discharge device can be screwed onto the internal thread of the rupture disk in order to drain off any escaping medium RUPTURE DISK i 1 2 3 A0029940 1 Rupture disk label 2 Rupture disk with 1 2 NPT internal thread with 1 width across flat 3 Transport protection For information on the dimensions see the Mechanical construction section accessories Wall mounting LWARNING...

Page 58: ...grommet Zero point adjustment All measuring devices are calibrated in accordance with state of the art technology Calibration takes place under reference conditions 49 Therefore a zero point adjustment in the field is generally not required Experience shows that zero point adjustment is advisable only in special cases To achieve maximum measuring accuracy even with low flow rates Under extreme pro...

Page 59: ... certificate option JP 50 to 60 C 58 to 140 F Readability of the local display 20 to 60 C 4 to 140 F The readability of the display may be impaired at temperatures outside the temperature range Dependency of ambient temperature on medium temperature 60 If operating outdoors Avoid direct sunlight particularly in warm climatic regions You can order a weather protection cover from Endress Hauser 104 ...

Page 60: ... 27 6 ms 50 g Shock resistance Shock due to rough handling following IEC 60068 2 31 Interior cleaning Cleaning in place CIP Sterilization in place SIP Options Oil and grease free version for wetted parts without declaration Order code for Service option HA Electromagnetic compatibility EMC As per IEC EN 61326 and NAMUR Recommendation 21 NE 21 Device version with PROFIBUS DP Complies with emission ...

Page 61: ...m Cubemass C 500 digital 60 C 140 F 205 C 401 F 60 C 140 F 90 C 194 F 25 C 77 F 205 C 401 F Cubemass C 500 60 C 140 F 205 C 401 F 60 C 140 F 160 C 320 F 55 C 131 F 205 C 401 F Seals For mounting sets with screwed on connections Viton 15 to 200 C 5 to 392 F EPDM 40 to 160 C 40 to 320 F Silicon 60 to 200 C 76 to 392 F Kalrez 20 to 275 C 4 to 527 F Density 0 to 5000 kg m3 0 to 312 lb cf Pressure temp...

Page 62: ...400 0 10 20 30 40 50 bar psi 200 400 600 0 Class 300 Class 150 A0027771 EN 30 With flange material 1 4539 904L lap joint flanges not wetted 1 4404 316 316L Flange connection according to JIS B2220 Order code for Mounting kit option PH 10K 300 F 350 250 200 150 100 50 0 50 C 200 150 100 50 0 50 400 0 10 20 bar psi 200 0 A0027778 EN 31 With flange material 1 4539 904L lap joint flanges not wetted 1 ...

Page 63: ...s a need to drain the leaking medium into a discharge device the sensor should be fitted with a rupture disk Connect the discharge to the additional threaded connection 72 If the sensor is to be purged with gas gas detection it should be equipped with purge connections Do not open the purge connections unless the containment can be filled immediately with a dry inert gas Use only low pressure to p...

Page 64: ...aximum full scale value In most applications 20 to 50 of the maximum full scale value can be considered ideal A low full scale value must be selected for abrasive media such as liquids with entrained solids flow velocity 1 m s 3 ft s For gas measurement the following rules apply The flow velocity in the measuring tubes should not exceed half the sound velocity 0 5 Mach The maximum mass flow depend...

Page 65: ...ot water or steam Via heating jackets NOTICE Danger of overheating when heating Ensure that the temperature at the lower end of the transmitter housing does not exceed 80 C 176 F Ensure that sufficient convection takes place at the transmitter neck Ensure that a sufficiently large area of the transmitted neck remains exposed The uncovered part serves as a radiator and protects the electronics from...

Page 66: ...0 187 24 21 Order code for Transmitter housing option D Polycarbonate and order code for Integrated ISEM electronics option A Sensor A mm F mm G mm N mm P mm Q mm 177 234 90 197 17 22 Housing of Proline 500 transmitter Hazardous area Zone 2 Class I Division 2 or Zone 1 Class I Division 1 T G A B C Q Q F A0033788 Order code for Transmitter housing option A Aluminum coated and order code for Integra...

Page 67: ...2 22 10 40 175 2 30 2 120 52 22 10 40 175 4 30 3 9 120 52 22 10 40 175 6 30 5 35 120 52 22 10 40 175 Order code for Sensor connection housing option A Aluminum coated DN A 1 B 1 C E F G mm mm mm mm mm mm mm 1 to 6 148 94 54 162 214 136 1 Depending on the cable gland used values up to 30 in Order code for Sensor connection housing option B Stainless hygienic DN A 1 B C E F G mm mm mm mm mm mm mm 1 ...

Page 68: ...mm mm mm 1 to 6 124 68 56 157 209 112 1 Depending on the cable gland used values up to 30 in Order code for Sensor connection housing option L Cast stainless DN A 1 B C E F G mm mm mm mm mm mm mm 1 to 6 145 86 59 185 237 136 1 Depending on the cable gland used values up to 30 in Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 69: ... mm 1 5 2 0 4 VCO 4 1 4539 904L order code for Process connection option HAW DN mm A in B mm L mm 1 SW 12 5 175 2 SW 12 5 175 4 SW 12 5 175 8 VCO 4 1 4404 316 316L order code for Process connection option CVS DN mm A in B mm L mm 6 AF 1 20 175 Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 70: ...L order code for Accessories option PF DN mm A mm B mm C mm D mm E mm L mm 1 to 6 90 0 60 3 4 Ø15 7 17 7 15 7 278 DN 1 to 4 with 4 VCO 4 DN 6 with 8 VCO 4 Lap joint flanges not wetted made of stainless steel 1 4404 316 316L Sealing sets order code for Enclosed accessories option P1 Viton P2 EPDM P3 silicone P4 Kalrez Flange according to ASME B16 5 Class 300 1 4539 904L order code for Accessories o...

Page 71: ... mm D mm E mm L mm 1 to 6 95 70 4 Ø15 28 15 0 278 DN 1 to 4 with 4 VCO 4 DN 6 with 8 VCO 4 Lap joint flanges not wetted made of stainless steel 1 4404 316 316L Sealing sets order code for Enclosed accessories option P1 Viton P2 EPDM P3 silicone P4 Kalrez Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 72: ... 4539 904L order code for Accessories option PD DN mm A in B in L mm 6 AF NPT 246 Sealing sets order code for Enclosed accessories option P1 Viton P2 EPDM P3 silicone P4 Kalrez Accessories Rupture disk purge connections SW 1 2 NPT 1 NPT 35 1 38 ca 75 approx 2 95 1 2 A0029960 1 Connection nipple for purge connections order code for Sensor options option CH Purge connection 2 Connection nipple with ...

Page 73: ...2 0 47 280 11 0 255 10 0 134 5 3 30 1 18 A0029553 35 Weather protection cover for Proline 500 External WLAN antenna The external WLAN antenna is not suitable for use in hygienic applications Proline 500 digital External WLAN antenna mounted on device 105 4 1 A0033607 36 Engineering unit mm in Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 74: ...0 2 8 1500 59 1 72 2 8 A0033606 37 Engineering unit mm in Proline 500 External WLAN antenna mounted on device 68 2 7 105 4 1 173 6 8 A0028923 38 Engineering unit mm in External WLAN antenna mounted with cable The external WLAN antenna can be mounted separately from the transmitter if the transmission reception conditions at the transmitter mounting location are poor Заказывайте на сайте www tinnov...

Page 75: ...uminum coated and order code for Integrated ISEM electronics option A Sensor A in F in G in N in P in Q in 6 57 9 13 3 15 7 36 0 94 0 83 Order code for Transmitter housing option D Polycarbonate and order code for Integrated ISEM electronics option A Sensor A in F in G in N in P in Q in 6 97 9 21 3 54 7 76 0 67 0 87 Housing of Proline 500 transmitter Hazardous area Zone 2 Class I Division 2 or Zon...

Page 76: ...on L Cast stainless and order code for Integrated ISEM electronics option B Transmitter A in B in C in F in G in Q in T in 7 40 3 35 4 06 11 6 8 54 5 12 9 41 Sensor connection housing A B C G D L E F S S P K Q N M A0030029 DN D K M N P Q S L in in in in in in in in in 1 18 0 05 4 72 2 05 0 87 0 39 1 57 6 89 1 18 0 08 4 72 2 05 0 87 0 39 1 57 6 89 1 18 0 15 4 72 2 05 0 87 0 39 1 57 6 89 1 18 0 211 ...

Page 77: ... 2 32 6 22 8 27 5 28 1 Depending on the cable gland used values up to 1 18 in Order code for Sensor connection housing option C Ultra compact hygienic stainless DN A 1 B C E F G in in in in in in in to 4 88 2 68 2 20 6 18 8 23 4 41 1 Depending on the cable gland used values up to 1 18 in Order code for Sensor connection housing option L Cast stainless DN A 1 B C E F G in in in in in in in to 5 71 ...

Page 78: ...inch 0 06 0 08 4 VCO 4 1 4539 904L order code for Process connection option HAW DN mm A in B in L in SW 0 49 6 89 SW 0 49 6 89 SW 0 49 6 89 8 VCO 4 1 4404 316 316L order code for Process connection option CVS DN mm A in B in L in AF 1 0 79 6 89 Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 79: ...d made of stainless steel 1 4404 316 316L Sealing sets order code for Enclosed accessories option P1 Viton P2 EPDM P3 silicone P4 Kalrez Flange according to ASME B16 5 Class 300 1 4539 904L order code for Accessories option PG DN in A in B in C in D in E in L in to 3 74 2 37 4 Ø0 62 0 81 0 62 10 94 DN to with 4 VCO 4 DN with 8 VCO 4 Lap joint flanges not wetted made of stainless steel 1 4404 316 3...

Page 80: ...1 4539 904L order code for Accessories option PD DN mm A in B in L in AF NPT 9 69 Sealing sets order code for Enclosed accessories option P1 Viton P2 EPDM P3 silicone P4 Kalrez Accessories Rupture disk purge connections SW 1 2 NPT 1 NPT 35 1 38 ca 75 approx 2 95 1 2 A0029960 1 Connection nipple for purge connections order code for Sensor options option CH Purge connection 2 Connection nipple with ...

Page 81: ...2 0 47 280 11 0 255 10 0 134 5 3 30 1 18 A0029553 41 Weather protection cover for Proline 500 External WLAN antenna The external WLAN antenna is not suitable for use in hygienic applications Proline 500 digital External WLAN antenna mounted on device 105 4 1 A0033607 42 Engineering unit mm in Заказывайте на сайте www tinnova ru endress_hauser Эл почта info tinnova ru ...

Page 82: ...0 2 8 1500 59 1 72 2 8 A0033606 43 Engineering unit mm in Proline 500 External WLAN antenna mounted on device 68 2 7 105 4 1 173 6 8 A0028923 44 Engineering unit mm in External WLAN antenna mounted with cable The external WLAN antenna can be mounted separately from the transmitter if the transmission reception conditions at the transmitter mounting location are poor Заказывайте на сайте www tinnov...

Page 83: ...ainless 3 7 kg 8 2 lbs Weight in SI units DN mm Weight kg 1 6 3 5 Weight in US units DN in Weight lbs to 8 Materials Transmitter housing Housing of Proline 500 digital transmitter Order code for Transmitter housing Option A Aluminum coated aluminum AlSi10Mg coated Option D Polycarbonate polycarbonate Housing of Proline 500 transmitter Order code for Transmitter housing Option A Aluminum coated alu...

Page 84: ... cable entry with internal thread G or NPT 4 Device plugs Cable entries and adapters Material Cable gland M20 1 5 Plastic Adapter for cable entry with internal thread G Adapter for cable entry with internal thread NPT Only available for certain device versions Order code for Transmitter housing Option A Aluminum coated Option D Polycarbonate Order code for Sensor connection housing Proline 500 dig...

Page 85: ...smitter PVC cable with copper shield Connecting cable for sensor Proline 500 transmitter Standard cable PVC cable with copper shield Reinforced cable PVC cable with copper shield and additional steel wire braided jacket UV rays can impair the cable outer sheath Protect the cable from exposure to sun as much as possible Sensor housing Acid and alkali resistant outer surface Stainless steel 1 4301 3...

Page 86: ...or user specific tasks Commissioning Operation Diagnostics Expert level Fast and safe commissioning Guided menus Make it run wizards for applications Menu guidance with brief descriptions of the individual parameter functions Device access via Web server or SmartBlue app 106 WLAN access to the device via mobile handheld terminal tablet or smart phone Reliable operation Operation in local language ...

Page 87: ... G 4 line illuminated graphic display touch control WLAN Information about WLAN interface 93 1 2 A0028232 47 Operation with touch control 1 Proline 500 digital 2 Proline 500 Display elements 4 line illuminated graphic display White background lighting switches to red in event of device errors Format for displaying measured variables and status variables can be individually configured Permitted amb...

Page 88: ... remote operation via HART protocol passive 1 Control system e g PLC 2 Transmitter power supply unit e g RN221N with communication resistor 3 Connection for Commubox FXA195 and Field Communicator 475 4 Field Communicator 475 5 Computer with Web browser e g Internet Explorer for access to the integrated device Web server or computer with an operating tool e g FieldCare DeviceCare AMS Device Manager...

Page 89: ... H1 network 7 Power supply FF H1 network 8 T box 9 Measuring device Via PROFIBUS DP network This communication interface is available in device versions with PROFIBUS DP 1 2 3 4 4 4 A0020903 51 Options for remote operation via PROFIBUS DP network 1 Automation system 2 Computer with PROFIBUS network card 3 PROFIBUS DP network 4 Measuring device Via PROFIBUS PA network This communication interface i...

Page 90: ...lable in device versions with a Modbus RS485 output 1 2 3 A0029437 53 Options for remote operation via Modbus RS485 protocol active 1 Control system e g PLC 2 Computer with Web browser e g Internet Explorer for accessing the integrated device Web server or with operating tool e g FieldCare DeviceCare with COM DTM CDI Communication TCP IP or Modbus DTM 3 Transmitter Via EtherNet IP network This com...

Page 91: ...is integrated via the terminal connection for signal transmission output 1 and the service interface CDI RJ45 4 2 1 3 5 5 5 A0033725 55 Options for remote operation via EtherNet IP network ring topology 1 Automation system e g RSLogix Rockwell Automation 2 Workstation for measuring device operation with Custom Add On Profile for RSLogix 5000 Rockwell Automation or with Electronic Data Sheet EDS 3 ...

Page 92: ... e g Scalance X204 Siemens 4 Measuring device Ring topology This communication interface is available in device versions with PROFINET 3 1 2 4 4 4 A0033719 57 Options for remote operation via PROFINET network ring topology 1 Automation system e g Simatic S7 Siemens 2 Computer with Web browser e g Internet Explorer for accessing the integrated device Web server or computer with operating tool e g F...

Page 93: ... for accessing the integrated device Web server or with FieldCare DeviceCare operating tool with COM DTM CDI Communication TCP IP or Modbus DTM 2 Standard Ethernet connecting cable with RJ45 connector 3 Service interface CDI RJ45 of the measuring device with access to the integrated Web server Proline 500 transmitter 2 3 1 A0027563 59 Connection via service interface CDI RJ45 1 Computer with Web b...

Page 94: ... IEEE 802 11i Configurable WLAN channels 1 to 11 Degree of protection IP67 Available antennas Internal antenna External antenna optional In the event of poor transmission reception conditions at the place of installation Available as an accessory 104 Only one antenna active in each case Max range 50 m 164 ft Materials External WLAN antenna Antenna ASA plastic acrylic ester styrene acrylonitrile an...

Page 95: ...l WLAN Special Documentation for the OPC UA Server application package 108 Supported operating tools Different operating tools can be used for local or remote access to the measuring device Depending on the operating tool used access is possible with different operating units and via a variety of interfaces Supported operating tools Operating unit Interface Additional information Web browser Noteb...

Page 96: ...the user to monitor the status of the device Furthermore the device data can be managed and the network parameters can be configured A device that has a WLAN interface can be ordered as an option is required for the WLAN connection order code for Display operation option G 4 line illuminated touch control WLAN The device acts as an Access Point and enables communication by computer or a mobile han...

Page 97: ...on again immediately without any errors If exchanging the electronics module e g I O electronics module Once the electronics module has been replaced the software of the module is compared against the current device firmware The module software is upgraded or downgraded where necessary The electronics module is available for use immediately afterwards and no compatibility problems occur Manual Add...

Page 98: ...oval The measuring device is certified for use in hazardous areas and the relevant safety instructions are provided in the separate Safety Instructions XA document Reference is made to this document on the nameplate The separate Ex documentation XA containing all the relevant explosion protection data is available from your Endress Hauser sales center Proline 500 digital ATEX IECEx Currently the f...

Page 99: ...2 AEx Ex nA IIC T5 T1 Gc Ex tb Transmitter Sensor AEx Ex ia IIIC Zone 21 AEx Ex ia tb IIIC T C Db Proline 500 ATEX IECEx Currently the following versions for use in hazardous areas are available Ex db eb Transmitter Sensor Category Type of protection Category Type of protection II2G Ex db eb ia IIC T6 T4 Gb II1 2G Ex ia IIC T6 T1 Ga Gb II2G Ex db eb ia IIC T6 T4 Gb II2G Ex ia IIC T6 T1 Gb Ex db Tr...

Page 100: ... T4 Gb Class I Zone 1 AEx Ex ia IIC T6 T1 Ga Gb Class I Zone 1 AEx Ex d ia IIC T6 T4 Gb Class I Zone 1 AEx Ex ia IIC T6 T1 Gb Ex nA Transmitter Sensor Class I Zone 2 AEx Ex nA IIC T5 T4 Gc Class I Zone 2 AEx Ex nA IIC T5 T1 Gc Ex tb Transmitter Sensor Zone 21 AEx Ex tb IIIC T85 C Db Zone 21 AEx Ex ia tb IIIC T C Db Pharmaceutical compatibility FDA USP Class VI TSE BSE Certificate of Suitability Fu...

Page 101: ...o be operated with certified devices of other manufacturers interoperability EtherNet IP certification The measuring device is certified and registered by the ODVA Open Device Vendor Association The measuring system meets all the requirements of the following specifications Certified in accordance with the ODVA Conformance Test EtherNet IP Performance Test EtherNet IP PlugFest compliance The devic...

Page 102: ...ation Detailed ordering information is available from the following sources In the Product Configurator on the Endress Hauser website www endress com Click Corporate Select your country Click Products Select the product using the filters and search field Open product page The Configure button to the right of the product image opens the Product Configurator From your Endress Hauser Sales Center www...

Page 103: ...tem for the purpose of preventive maintenance or process analysis These data enable the operator to Draw conclusions using these data and other information about the impact process influences such as corrosion abrasion buildup etc have on the measuring performance over time Schedule servicing in time Monitor the process or product quality e g gas pockets Concentration Package Description Concentra...

Page 104: ...nsmitter Installation Instructions EA01151 Proline 500 transmitter Installation Instructions EA01152 External WLAN antenna External WLAN antenna with 1 5 m 59 1 in connecting cable and two angle brackets Order code for Enclosed accessories option P8 Wireless antenna wide area The external WLAN antenna is not suitable for use in hygienic applications Further information on the WLAN interface 93 Ord...

Page 105: ...nitoring of connected 4 20 mA measuring devices via a Web browser Technical Information TI00025S Operating Instructions BA00053S Fieldgate FXA520 Gateway for the remote diagnostics and remote configuration of connected HART measuring devices via a Web browser Technical Information TI00025S Operating Instructions BA00051S Field Xpert SFX350 Field Xpert SFX350 is a mobile computer for commissioning ...

Page 106: ...ant asset management tool from Endress Hauser It can configure all smart field units in your system and helps you manage them By using the status information it is also a simple but effective way of checking their status and condition Operating Instructions BA00027S and BA00059S DeviceCare Tool to connect and configure Endress Hauser field devices Innovation brochure IN01047S System components Acc...

Page 107: ...D KA01351D Proline 500 KA01314D KA01291D KA01389D KA01318D KA01347D KA01350D Operating Instructions Measuring device Documentation code HART FOUNDATION Fieldbus PROFIBUS PA PROFIBUS DP Modbus RS485 EtherNet IP PROFINET Cubemass C 500 BA01527 D BA01560D BA01549D BA01871D BA01538D BA01748D BA01759D Description of Device Parameters Measuring device Documentation code HART FOUNDATION Fieldbus PROFIBUS...

Page 108: ...D SD02205D SD01701D SD01991D SD01985D Concentration measurement SD01716D SD01717D SD01719D SD02215D SD01718D SD02011D SD02010D Installation Instructions Contents Comment Installation instructions for spare part sets and accessories Documentation code specified for each individual accessory Registered trademarks HART Registered trademark of the FieldComm Group Austin Texas USA PROFIBUS Registered t...

Reviews: