Optional Items

PXI slot blocker kit (NI part number 199198-01)

SCB-19 connector block (NI part number 785642-01)

SHH19-MH19-AUX shielded single-ended cable, 1 M (NI part number 784091-01)

Visit 

ni.com

 for more information about these additional items.

Preparing the Environment

Ensure that the environment you are using the PXIe-5413 in meets the following
specifications:

Maximum altitude

2,000 m (800 mbar) (at 25 °C ambient
temperature)

Pollution Degree

2

Operating ambient temperature range

0 °C to 55 °C (Tested in accordance with
IEC 60068-2-1 and IEC 60068-2-2. Meets
MIL-PRF-28800F Class 3 low temperature
limit and MIL-PRF-28800F Class 2 high
temperature limit.)

Operating relative humidity range

10% to 90%, noncondensing (Tested in
accordance with IEC 60068-2-56.)

Indoor use only.

Related Information

Refer to the PXIe-5413 Specifications at ni.com for complete specifications.

Installing the Software

You must be an Administrator to install NI software on your computer.
1.

Install an ADE, such as LabVIEW or LabWindows

/CVI

.

2.

Visit 

ni.com/downloads/drivers

 and search for 

NI-FGEN

.

3.

Download the latest version of NI-FGEN by clicking 

Download

 and then running the

executable (

.exe

) file.

4.

Follow the instructions in the installation prompts.

Note

 

Windows users may see access and security messages during

installation. Accept the prompts to complete the installation.

5.

When the installer completes, select 

Restart

 in the dialog box that prompts you to restart,

shut down, or restart later.

4

  | 

ni.com

  | 

PXIe-5413 Getting Started Guide

Summary of Contents for PXIe-5413

Page 1: ...Electromagnetic Compatibility Guidelines 2 Verifying the System Requirements 2 Unpacking the Kit 2 Other Equipment 3 Preparing the Environment 4 Installing the Software 4 Installing the PXIe 5413 5 PX...

Page 2: ...product in strict accordance with the instructions in the product documentation Furthermore any changes or modifications to the product not expressly approved by National Instruments could void your a...

Page 3: ...xposed pins of connectors Note Do not install a module if it appears damaged in any way 3 Unpack any other items and documentation from the kit Store the module in the antistatic package when the modu...

Page 4: ...Operating relative humidity range 10 to 90 noncondensing Tested in accordance with IEC 60068 2 56 Indoor use only Related Information Refer to the PXIe 5413 Specifications at ni com for complete spec...

Page 5: ...s so that inlet and outlet vents are not obstructed For more information about optimal chassis positioning refer to the chassis documentation 5 Remove the black plastic covers from all the captive scr...

Page 6: ...ectrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 12 Cover all empty slots using EMC filler panels or fill using slot blockers to...

Page 7: ...alog output terminal 2 CH 1 3 PFI 0 Input Output Imports digital trigger signals and exports digital event signals Imported digital trigger signals can start or step through waveform generation and ex...

Page 8: ...eset does not show an error The PXIe 5413 has powered down because the internal temperature exceeded the maximum limit The over temperature condition must be corrected and the hardware reset To reset...

Page 9: ...ce for signals 5 NC No connection 6 GND Ground reference for signals 7 GND Ground reference for signals 8 AUX 0 PFI 0 Bidirectional PFI line 9 AUX 0 PFI 1 Bidirectional PFI line 10 GND Ground referenc...

Page 10: ...igital signals to the AUX 0 connector on the PXIe 5413 front panel Refer to the following figure and table for information about the SCB 19 signals when connected to the AUX 0 front panel connector 1...

Page 11: ...appear under the name of their associated chassis 3 Expand your Chassis tree item MAX lists all modules installed in the chassis Your default names may vary Note If you do not see your module listed p...

Page 12: ...ule that is in the same chassis as the PXIe 5413 is installed uninstalled or moved When the PXIe 5413 is in an environment where the ambient temperature varies or the PXIe 5413 temperature has drifted...

Page 13: ...oad dialog box choose a waveform from the Waveform list 4 Click OK 5 Click Run to start waveform generation To optimize the waveform select Edit Device Configuration and adjust the settings NI FGEN In...

Page 14: ...D fails to light after you power on the chassis a problem may exist with the chassis power rails a hardware module or the LED Caution Apply external signals only while the PXIe 5413 is powered on Appl...

Page 15: ...Device Manager a Under an NI entry confirm that a PXIe 5413 entry appears Note If you are using a PC with a device for PXI remote control system under System Devices also confirm that no error conditi...

Page 16: ...16 Bit AWG NI PXI 5413 PXIe 5413 Firmware Licensing Software is licensed pursuant to the software license agreements provided with the software or in the absence of such license agreements the Nationa...

Page 17: ...fication If your product supports calibration you can obtain the calibration certificate for your product at ni com calibration NI corporate headquarters is located at 11500 North Mopac Expressway Aus...

Page 18: ...ents EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy a...

Reviews: