Step 5. Confirm Basic Module Operation
Purpose of Confirmation
The purpose of this process is to confirm the basic module operation of
the PXI-1430B.
Confirmation Requirements
The PXI-1430B is confirmed by measuring the IF center frequency and
bandwidth with a 1.5 GHz input frequency. This process confirms that
all connectors, switches, cables, and related circuitry used in these
signal paths are functional.
Equipment Requirements
shows a block diagram of the equipment setup with the type
of equipment (or equivalent performance) required to confirm basic
module operation.
Table 5
Equipment Requirements to Confirm Basic Module Operation
Qty
Equipment
Description
1
Agilent 83650A
Signal Generator, 26.5 GHz
1
Agilent 8566B
Spectrum Analyzer
1
NI PXIe-1065
PXI chassis
1
NI PXI-8106
PXI Embedded Controller with
Windows 7/XP
1
Display