ni.com

4. If you need to use an EMI gasket to reduce high-frequency emissions, 

install it now. Refer to the 

Retrofitting Your PXI Module Note to Users

 

for gasket installation instructions.

5. Remove the packaging material on the PXI connector and on the screws.

6. Ensure that the ejector handle is in the unlatched (downward) position, 

as shown in Figure 1.

7. Holding the module by the ejector handle, slide it into an empty slot, 

ensuring that the base card (on the left when looking at the front of the 
module) engages with the card guides in the chassis.

8. Slide the module completely into the chassis and latch by pulling up on 

the ejector handle.

9. Tighten the captive screws at the top and bottom of the module front panel.

10. Before operating the module, install all chassis covers and filler panels. 

Missing filler panels disrupt the necessary air circulation in the chassis.

Note

NI recommends that you install slot blockers between 

modules to maximize air flow. NI recommends using the PXI 
Chassis Slot Blocker Kit, part number 778678-01, available for 
purchase at 

ni.com

.

11. Plug in and power on the chassis.

NI PXI modules are sensitive instruments that should be handled carefully. 
Do not expose the module to temperatures or humidity beyond the rated 
maximums. Keep the module free of dust by cleaning with compressed air 
only. Do not clean the module with any solvents or liquids. For more 
information about module cooling, refer to the guidelines in the 

Maintain 

Forced-Air Cooling Note to Users

 included with your NI digital waveform 

generator/analyzer.

Maintaining PXI Systems

Clean the fan filters on the chassis regularly to prevent air circulation path 
blockage. Cleaning frequency depends on the amount of use and the 
operating environment. For specific information about cleaning procedures, 
and other recommended maintenance, refer to the module specifications and 
the chassis user documentation.

Uninstalling PXI Modules

When removing PXI modules from the chassis, first power down the chassis. 
Then ensure that you are grounded with a grounding strap or are touching a 
grounded metal surface. To avoid ESD, do not touch the exposed pins of the 
PXI connector or any exposed circuitry on the module. When not in use, PXI 
modules should be stored in the original antistatic packaging to avoid 
damage. 

Summary of Contents for NI

Page 1: ...Digital Waveform Generator Analyzer Help at Start Programs National Instruments NI HSDIO Documentation NI Digital Waveform Generator Analyzer Help Refer to the specifications document that ships with...

Page 2: ...n ellipsis represent a range of values associated with a bit or signal name for example AO 3 0 The symbol leads you through nested menu items and dialog box options to a final action The sequence File...

Page 3: ...ames of disk drives paths directories programs subprograms subroutines device names functions operations variables filenames and extensions monospace bold Bold text in this font denotes the messages a...

Page 4: ...uch the exposed pins of connectors To avoid ESD damage in handling the device take the following precautions Ground yourself with a grounding strap or by touching a grounded object Touch the antistati...

Page 5: ...sket Refer to step 5 Installing the Hardware for more information about the gasket 4 Installing the Software This section describes the software installation process for the NI digital waveform genera...

Page 6: ...appears that asks if you want to restart shut down or restart later Select Restart 4 If you are using a system running the LabVIEW Real Time Module download NI HSDIO to the target using MAX Refer to...

Page 7: ...out re enabling your device refer to the I Received a Thermal Shutdown Error section in Appendix B Troubleshooting Installing a PXI Module To install the module refer to Figure 1 and complete the foll...

Page 8: ...78678 01 available for purchase at ni com 11 Plug in and power on the chassis NI PXI modules are sensitive instruments that should be handled carefully Do not expose the module to temperatures or humi...

Page 9: ...vice To install your PCI device complete the following steps 1 Power off and unplug the PC 2 If the PC has multiple fan speed settings ensure that the fans are set to the highest setting 3 Remove the...

Page 10: ...uring and Testing in MAX To configure and test the NI digital waveform generator analyzer in MAX complete the following steps 1 Launch MAX to configure and test the digital waveform generator analyzer...

Page 11: ...e digital device You need this information when you program the device Using the Test Panel to Generate and Acquire Data To verify your device configuration use the device test panel in MAX to generat...

Page 12: ...in the following figure Because the data channels on the NI digital waveform generator analyzer are bidirectional while the data is being continuously generated on all of the channels as in the previo...

Page 13: ...tor analyzer Note Whether you use NI cables and accessories or design your own you should properly terminate cables to avoid improper measurements due to signal reflections overshoot and undershoot Re...

Page 14: ...erator Analyzer 2 NI SHC68 C68 D2 Cable 3 NI CB 2162 Accessory 1 PXI Chassis with NI Digital Waveform Generator Analyzer 2 NI SMB 2163 Accessory 3 NI SHC68 C68 D2 Cable NI PXI 6551 50 MHz Digital I O...

Page 15: ...s to various types of devices and circuits for interfacing testing or analysis This cable offers connectivity similar to that found on a typical logic analyzer so you can use it in logic analyzer type...

Page 16: ...his cable offers connectivity similar to that found on a typical logic analyzer so you can use it in logic analyzer type applications The cable is shown in Figure 8 1 NI Digital Waveform Generator Ana...

Page 17: ...ctor for the VHDCI cable from NI part number 778914 01 For more information about creating these custom accessories refer to the Interfacing to the NI Digital Waveform Generator Analyzer using the VHD...

Page 18: ...ation an NI digital waveform generator analyzer generates data through a matched impedance system that consists of a 50 output impedance a 50 cable and a 50 accessory Figure 9 shows a diagram of such...

Page 19: ...operations and attributes that exercise all the functionality of the device including configuration control and other device specific functions Information about programming with NI HSDIO is availabl...

Page 20: ...or your first NI HSDIO program select to Start All Programs Program Files NI HSDIO Examples NI Example Finder You can also use the NI Example Finder to search or browse examples NI HSDIO examples are...

Page 21: ...3 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 DIO 31 GND DIO 29 GND DIO 27 G...

Page 22: ...3 45 47 49 51 53 55 57 59 61 63 65 DIO 0 31 Data Bidirectional digital I O data channels 0 through 31 26 30 64 PFI 1 3 Control Input terminals to the NI 654x for external triggers or output terminals...

Page 23: ...25 26 27 28 29 30 31 32 33 34 RESERVED GND RESERVED GND RESERVED GND RESERVED RESERVED RESERVED GND RESERVED GND DIO 19 GND DIO 17 GND DIO 15 GND DIO 13 GND DIO 11 GND DIO 9 GND DIO 7 PFI 1 DIO 5 GND...

Page 24: ...s those described in Tables 5 and 6 Table 2 NI 655x DDC Connector Pins Pins Signal Name Signal Type Signal Description 33 DDC CLKOUT Control Output terminal for the exported Sample clock 67 STROBE Con...

Page 25: ...11 GND DIO 12 GND DIO 13 GND DIO 14 GND DIO 15 GND STROBE GND DDC CLK OUT LVDS GND RESERVED NC GND DDC CLK OUT LVPECL GND PFI_1 PFI_2 PFI_3 GND RESERVED NC DIO 0 DIO 1 DIO 2 DIO 3 DIO 4 DIO 5 DIO 6 D...

Page 26: ...Complementary external Sample clock source which can be used for dynamic acquisition 14 17 20 23 26 29 32 35 38 41 44 47 50 53 56 59 DIO 0 15 Data Bidirectional digital I O data channels 0 through 15...

Page 27: ...O 2 GND DIO 3 GND DIO 4 GND DIO 5 GND DIO 6 GND DIO 7 GND DIO 8 GND DIO 9 GND DIO 10 GND DIO 11 GND DIO 12 GND DIO 13 GND DIO 14 GND DIO 15 GND STROBE GND DDC CLK OUT LVDS GND RESERVED NC GND DDC CLK...

Page 28: ...gers or output terminal from the device for events CLK OUT Reference Clock Output Control Terminal for the exported PLL Reference clock or the exported Sample clock Table 5 ACTIVE LED Indicators Color...

Page 29: ...damaged device 4 Reinstall the PXI module using the procedure described in step 5 Installing the Hardware 5 Power on your PXI chassis 6 Verify the device appears in MAX 7 Reset the device in MAX and p...

Page 30: ...hat your device is effectively cooled 3 Call niHSDIO_ResetDevice NI HSDIO Reset Device VI or perform a device reset in MAX For more information about performing a device reset in MAX refer to step 6 C...

Page 31: ...cated at 11500 North Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world to help address your support needs For telephone support in the United Stat...

Page 32: ...National Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer t...

Reviews: