Chapter 2
Using the SCXI-1129
2-16
ni.com
Matrix Expansion
In addition to the configurations discussed above, it is possible to combine
modules with the same configuration to create much larger matrixes. You
can expand matrix modules through the terminal blocks by connecting
columns and/or rows. Figure 2-11 shows an example of an 8
× 64
matrix
created by connecting two 8
× 32
matrixes using a matrix expansion cable.
Figure 2-11.
8
×
64 Matrix Schematic
•
•
•
•
•
•
Cab
le
#1
Cab
le
#2
Terminal
Block
Boundary
SCXI-1129/SCXI-1336
SCXI-1129/SCXI-1336
C0
C1
C30
C31
C32 (C0)
R0
R1
R2
R3
R4
R5
R6
R7
C63 (C31)
Matrix
Expansion
Plug
Summary of Contents for NI SCXI-1336
Page 1: ...SCXI 1336...