background image

6.

Click 

Run

 to run the example program.

Setting Up SMC-Based Devices for
Synchronization

The following steps are required for any type of synchronization involving an SMC-based
device, including NI-TClk synchronization.

Note

 

If you plan to share triggers and/or clocks for the purpose of synchronizing

SMC-based devices, you must identify or configure certain components in MAX.

PXI Modules

You must identify the PXI system controller by completing the following steps:
1.

Launch MAX.

2.

In the MAX configuration pane, expand the chassis tree.

3.

Right-click the name of the chassis you are using, and select 

Identify As

.

4.

Select your controller from the list. For example, select 

External PC

 if you are using an

MXI controller in an external PC.

PXI Express Modules

Your device should automatically be configured for synchronization. If your device is not
automatically configured, contact NI Technical Support at 

ni.com/support

.

PCI Devices

You must configure the RTSI cable by completing the following steps:
1.

Connect an RTSI cable between the PCI devices to physically share triggers and/or
clocks.

2.

Launch MAX.

3.

In the MAX configuration pane, right-click 

Devices and Interfaces

, and select 

Create

New

.

4.

In the dialog box that opens, select 

NI-RTSI Cable

.

5.

Click 

Finish

.

6.

In the MAX configuration pane, right-click the NI-RTSI cable, select 

Add Device to NI-

RTSI Cable

, and select the device you want to add.

For more information about connecting PCI devices, refer to 

NI High-Speed Digitizers Help

»

Programming

»

Reference

»

NI-TClk Synchronization Help

.

12

  | 

ni.com

  | 

NI PXI/PXIe/PCI-5105 Getting Started Guide

Summary of Contents for NI PXIe-5105

Page 1: ...umentation Contents Safety and Electromagnetic Compatibility 2 Verifying the System Requirements 2 Unpacking the Kit 2 Preparing the Environment 3 PXI Express Modules 3 PXI Modules 3 PCI Modules 3 Kit...

Page 2: ...e the NI 5105 your system must meet certain requirements For more information about minimum system requirements recommended system requirements and supported ADEs refer to the readme for your selected...

Page 3: ...000 m 800 mbar at 25 C ambient temperature Pollution Degree 2 Indoor use only PXI Modules Operating Environment Ambient temperature range 0 C to 55 C Tested in accordance with IEC 60068 2 1 and IEC 6...

Page 4: ...fer to specifications document for your device at ni com manuals Kit Contents Figure 1 NI 5105 Kit Contents 1 2 3 5 6 7 8 4 1 NI PCI 5105 Device 2 NI PXI 5105 Device 3 NI PXIe 5105 Device 4 Screw Kit...

Page 5: ...er should open automatically If the installation window does not appear navigate to the drive double click it and double click autorun exe 3 Follow the instructions in the installation prompts Note Wi...

Page 6: ...dules can be placed in PXI Express peripheral slots PXI Express Hybrid peripheral slots or PXI Express System Timing slots 6 Touch any metal part of the chassis to discharge static electricity 7 Ensur...

Page 7: ...s can be inserted only into PCI slots 4 Touch any metal part of the computer to discharge any static electricity 5 Insert the module into the slot you selected Gently rock the module in to place witho...

Page 8: ...oscope CH 0 CH 5 CH 4 CH 3 CH 2 CH 1 PFI 1 CH 7 CH 6 60MHz Oscilloscope PFI 1 CH 0 CH 1 CH 2 CH 3 CH 4 CH 5 CH 6 CH 7 Table 1 NI 5105 Front Panel Signal Connections Connector Description Function CH 0...

Page 9: ...his identifier when programming the NI 5105 5 The MAX self test performs a basic verification of hardware resources To self test a module in MAX right click the module and select Self Test 6 Run the t...

Page 10: ...OPE examples directory where NIDocDir is one of the following directories Windows 8 7 Vista Users Public Documents National Instruments Windows XP Documents and Settings All Users Shared Documents Nat...

Page 11: ...to input signal PFI 1 using an SMB cable 2 Launch the NI SCOPE SFP from the Start menu 3 In the Select Device dialog box select the device name assigned to the device in MAX 4 To enable signal output...

Page 12: ...sing an MXI controller in an external PC PXI Express Modules Your device should automatically be configured for synchronization If your device is not automatically configured contact NI Technical Supp...

Page 13: ...that a National Instruments entry appears in the system device list Reinstall NI SCOPE and the device if error conditions appear in the list If you are using a MXI controller right click PCI to PCI B...

Page 14: ...oft Front Panel NI SCOPE Instrument Driver about the capabilities of your hardware with device specifications more about your products through ni com NI PXI PXIe PCI 5105 Specifications NI High Speed...

Page 15: ...ration of conformity This system affords the user protection for electromagnetic compatibility EMC and product safety You can obtain the DoC for your product by visiting ni com certification If your p...

Page 16: ...s and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance...