Chapter 1
Getting Started
NI PXIe-1071 User Manual
1-10
ni.com
An independent buffer drives PXIe_CLK100 to each peripheral slot. These
clocks are matched in skew to less than 100 ps. The differential pair must
be terminated on the peripheral with LVPECL termination for the buffer to
drive PXIe_CLK100 so that when there is no peripheral or a peripheral that
does not connect to PXIe_CLK100, there is no clock being driven on the
pair to that slot.
An independent buffer drives PXIe_SYNC100 to each peripheral slot. The
differential pair must be terminated on the peripheral with LVPECL
termination for the buffer to drive PXIe_SYNC100 so that when there is no
peripheral or a peripheral that does not connect to PXIe_SYNC100, there
is no clock being driven on the pair to that slot.
PXI_CLK10, PXIe_CLK100 and PXIe_SYNC100 have the default timing
relationship described in Figure 1-6.
Figure 1-6.
System Reference Clock Default Behavior
PXIe_CLK100
PXI_CLK10
PXIe_
S
YNC100
0 1 2
3
4 5 6 7
8
9 0 1 2
3
4 5 6 7
8
9 0 1 2
3
4 5 6 7
8
9