Appendix B
Connecting I/O Signals
©
National Instruments Corporation
B-5
Connecting to 5B and SSR Signal Conditioning
NI provides cables that allow you to connect signals from the
NI PXI-7831R directly to 5B backplanes for analog signal conditioning
and SSR backplanes for digital signal conditioning.
Table B-2.
NI PXI-7831R I/O Signal Summary
Signal Name
Driver
Type
Signal
Type and
Direction
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA at V)
Rise
Time
Bias
+5V
—
DO
—
—
—
—
—
—
AI<0..7>+
—
AI
10 G
Ω
in
parallel
with
100 pF
42/35
—
—
—
±2 nA
AI<0..7>–
—
AI
10 G
Ω
in
parallel
with
100 pF
42/35
—
—
—
±2 nA
AIGND
—
AO
—
—
—
—
—
—
AISENSE
—
AI
10 G
Ω
in
parallel
with
100 pF
42/35
—
—
—
±2 nA
AO<0..7>
—
AO
1.25
Ω
Short-
circuit to
ground
5 at 10
5 at –10
10 V/µs
—
AOGND
—
AO
—
—
—
—
—
—
DGND
—
DO
—
—
—
—
—
—
DIO<0..15>
Connector 0
DIO<0..39>
Connector<1..2>
Default
DIO
—
–0.5
to +7.0
5.4 at 2.4
5.0 at 0.4
12 ns
—
Slow
DIO
—
–0.5
to +7.0
1.9 at 0.4
1.9 at 0.4
75 ns
—
Fast
DIO
—
–0.5
to +7.0
16 at 2.4
14 at 0.4
6 ns
—
AI = Analog Input
AO = Analog Output
DIO = Digital Input/Output
DO = Digital Output