This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for
U.S. or 230 V for Europe.
Measurement Category III is for measurements performed in the
building installation at the distribution level. This category refers
to measurements on hard-wired hardware such as hardware in
fixed installations, distribution boards, and circuit breakers. Other
examples are wiring, including cables, bus bars, junction boxes,
switches, socket outlets in the fixed installation, and stationary
motors with permanent connections to fixed installations.
Caution
Do not connect the NI 9244 to signals or use
for measurements within Measurement Category IV.
Safety Guidelines for Hazardous Voltages
If hazardous voltages are connected to the device, take the
following precautions. A hazardous voltage is a voltage greater
than 42.4 Vpk voltage or 60 VDC to earth ground.
Caution
Ensure that hazardous voltage wiring is
performed only by qualified personnel adhering to local
electrical standards.