©
National Instruments Corporation
4-1
4
Analog Input
Figure 4-1 shows the analog input circuitry of M Series devices.
Figure 4-1.
M Series Analog Input Circuitry
The main blocks featured in the M Series analog input circuitry are as
follows:
•
I/O Connector
—You can connect analog input signals to the M Series
device through the I/O connector. The proper way to connect analog
input signals depends on the analog input ground-reference settings,
described in the
Analog Input Ground-Reference Settings
section. Also
refer to Appendix A,
, for device I/O
connector pinouts.
•
MUX
—Each M Series device has one analog-to-digital converter
(ADC). The multiplexers (MUX) route one AI channel at a time to the
ADC through the NI-PGIA.
•
Ground-Reference Settings
—The analog input ground-reference
settings circuitry selects between differential, referenced single-ended,
and non-referenced single-ended input modes. Each AI channel can
use a different mode.
•
Instrumentation Amplifier (NI-PGIA)
—The NI programmable gain
instrumentation amplifier (NI-PGIA) is a measurement and instrument
class amplifier that minimizes settling times for all input ranges. The
DIFF, R
S
E,
or NR
S
E
I/O Connector
AI <0..
n
>
M
u
x
AI
S
EN
S
E
AI GND
NI-PGIA
AI Termin
a
l
Config
u
r
a
tion
S
election
Inp
u
t R
a
nge
S
election
ADC
AI FIFO
AI D
a
t
a
AI Lowp
ass
Filter