National Instruments NI-5761 User Manual And Specifications Download Page 24

©

 National Instruments Corporation

23

NI 5761R User Guide and Specifications

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for electrical equipment for 
measurement, control, and laboratory use:

EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity

EN 55011 (CISPR 11): Group 1, Class A emissions

AS/NZS CISPR 11: Group 1, Class A emissions

FCC 47 CFR Part 15B: Class A emissions

ICES-001: Class A emissions

Note

For EMC declarations and certifications, refer to the 

Online Product Certification

 section of 

this document.

Caution

For EMC compliance, operate this device with shielded cables and accessories.

CE Compliance

This product meets the essential requirements of applicable European Directives as follows:

2006/95/EC; Low-Voltage Directive (safety)

2004/108/EC; Electromagnetic Compatibility Directive (EMC)

Online Product Certification

To obtain product certifications and the Declaration of Conformity for this product, visit 

ni.com/

certification

, search by model number or product line, and click the appropriate link in the 

Certification column.

Environmental Management

National Instruments is committed to designing and manufacturing products in an environmentally 
responsible manner. NI recognizes that eliminating certain hazardous substances from our products is 
beneficial not only to the environment but also to NI customers.

For additional environmental information, refer to the 

NI and the Environment

 Web page at 

ni.com/

environment

. This page contains the environmental regulations and directives with which NI 

complies, as well as other environmental information not included in this document.

Waste Electrical and Electronic Equipment (WEEE)

EU Customers

At the end of the product life cycle, all products 

must

 be sent to a WEEE recycling 

center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, 
and compliance with WEEE Directive 2002/96/EC on Waste and Electronic Equipment, 
visit

ni.com/environment/weee

.

⬉ᄤֵᙃѻક∵ᶧ᥻ࠊㅵ⧚ࡲ⊩ ˄Ё೑

RoHS

˅

Ё೑ᅶ᠋

National Instruments

ヺড়Ё೑⬉ᄤֵᙃѻકЁ䰤ࠊՓ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ

 (RoHS)

DŽ

݇Ѣ

National Instruments

Ё೑

RoHS

ড়㾘ᗻֵᙃˈ䇋ⱏᔩ

ni.com/environment/rohs_china

DŽ

(For information about China RoHS compliance, go to 

ni.com/environment/rohs_china

.)

Summary of Contents for NI-5761

Page 1: ...NI 5761...

Page 2: ...s Front Panel and Connector Pinouts 2 Block Diagram 4 NI 5761 Component Level Intellectual Property CLIP 5 Cables 6 Clocking 7 Using Your NI 5761R with a LabVIEW FPGA Example VI 7 Creating a LabVIEW P...

Page 3: ...before powering down the module and only connect signals after the adapter module has been powered on by the NI FlexRIO FPGA module Table 1 NI 5761 Front Panel Connectors Device Front Panel Connector...

Page 4: ...sis NI is not liable for any damage resulting from such signal connections For the maximum input and output ratings for each signal refer to the Specifications section of this document Table 2 NI 5761...

Page 5: ...nitialization Done Reinitialize Configuration Error Sample Clock Select Sample Clock Commit SPI Idle Analog FE IO Module Clock 0 n 1 Single Sample CLIP n 2 Multiple Sample CLIP Multiple Sample CLIP Si...

Page 6: ...IP integration functionality of the user defined CLIP but also allows the CLIP to communicate directly with circuitry external to the FPGA Adapter module socketed CLIP allows your IP to communicate di...

Page 7: ...e CLIP Generates one sample per clock cycle at a default sample rate of 250 MHz You can set a lower sample rate by using an external Sample clock This CLIP provides access to four analog input channel...

Page 8: ...I Note In NI application software NI FlexRIO adapter modules are referred to as IO Modules Complete the following steps to run an example that acquires a waveform on CH 0 of the NI 5761 1 Connect one...

Page 9: ...Channel box 9 Set the Trigger Level V and the Record Size controls to the desired value 10 In the Trigger Type box you can select either Software Trigger or Data Edge If you select Software Trigger th...

Page 10: ...25 MHz in the Compile for single frequency control Click OK 8 Right click the FPGA target and select New FPGA Base Clock again 9 In the Resource pull down menu select 200 MHz Clock Click OK 10 Right c...

Page 11: ...right click My Computer and select New VI A blank VI opens Select Window Show Block Diagram to open the VI block diagram 2 Add the Open FPGA VI Reference function located on the FPGA Interface palette...

Page 12: ...Reference function located on the FPGA Interface palette to the right of the While Loop on the block diagram 17 Wire the FPGA VI Reference Out indicator of the Read Write Control function to the FPGA...

Page 13: ...elp Embedded in LabVIEW Help Contains information about the basic functionality of LabVIEW FPGA Module NI FlexRIO Help Embedded in LabVIEW FPGA Module Help Contains FPGA module adapter module and CLIP...

Page 14: ...otherwise noted All graphs illustrate the performance of a representative module Typical values describe useful product performance that are not covered by warranty Typical values cover the expected p...

Page 15: ...dB 100 1 MHz 90 dB 501 MHz 80 dB AC Coupled Measurements Figure 8 AC Coupled Bandwidth Passband Table 5 AC Coupled Spectral Performance Measurement 20 17 MHz 70 17 MHz 123 17 MHz SNR 72 5 dB 71 4 dB 7...

Page 16: ...S Average Figure 10 AC Coupled Spectral Measurements 20 1 MHz 1 dBFS 8 192 Point FFT 10 RMS Average 20 0 115 110 105 100 95 90 85 80 75 70 65 60 55 50 45 40 35 30 25 20 15 10 5 0 Amplitude dBFS Freque...

Page 17: ...to 250 MHz Table 6 lists the DC coupled spectral performance measurements All values are measured with a 250 MHz external Sample clock Channel to channel isolation 1 MHz 90 dB 100 1 MHz 80 dB 501 MHz...

Page 18: ...you can add series resistance close to the source to properly bias the NI 5761 input terminals You can also use the NI 5761 input channel bias DACs to remove DC offset present in the system For more i...

Page 19: ...igure 14 DC Coupled Spectral Measurements 20 1 MHz 1 dBFS 8 192 Point FFT 10 RMS Average 0 115 110 105 100 95 90 85 80 75 70 65 60 55 50 45 40 35 30 25 20 0 Amplitude dBFS Frequency MHz 15 10 5 25 50...

Page 20: ...t FFT 10 RMS Average Analog Input Phase Noise Figure 16 Analog Input Phase Noise 0 115 110 105 100 95 90 85 80 75 70 65 60 55 50 45 40 35 30 25 20 0 Amplitude dBFS Frequency MHz 15 10 5 20 40 60 80 10...

Page 21: ...l Characteristics Number of channels 1 single ended Connector SMA Input impedance 50 Input coupling AC External Sample Clock Input voltage range 0 63 Vpk pk to 2 5 Vpk pk Input frequency range 175 MHz...

Page 22: ...out 50 Iout 2 mA Maximum toggle frequency 500 kHz Absolute maximum input 0 5 V to 7 V EEPROM Map Caution Only write to User Space Writing to any other offset may cause the NI 5761 to stop functioning...

Page 23: ...with IEC 60068 2 1 and IEC 60068 2 2 Relative humidity range 5 to 95 noncondensing tested in accordance with IEC 60068 2 56 Note Clean the device with a soft non metallic brush Make sure that the dev...

Page 24: ...ification search by model number or product line and click the appropriate link in the Certification column Environmental Management National Instruments is committed to designing and manufacturing pr...

Page 25: ...rth Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world to help address your support needs For telephone support in the United States create your se...

Reviews: