Tools needed:
•
2.5 mm flat-blade screwdriver (your own or NI part number 786911-01)
•
27 mm deep socket
•
30 mm socket or wrench
•
24 mm socket or wrench
•
(Optional) 24 mm gland dome nut tightening tool (SEALCON part number S-2400-WR)
Complete the following steps to connect power and ground.
1.
Confirm that the external power to the MON-10496 has been turned off.
2.
Install the PG16 gland on the device enclosure.
You may use the gland provided by NI or your own gland. Check the
MON-10496
Specifications
on
for the required gland specifications.
Note
If you are using an alternative gland or fitting, ensure that it is rated for
Type 3 or greater and install according to the manufacturer's instructions.
a)
Insert the PG16 gland through the hole on the enclosure.
b)
Install the backing nut on the PG16 gland and tighten the nut with a 30 mm socket or
wrench to a torque of 5 N · m (44.2 lb · in.) while holding the gland body with a
27 mm deep socket.
3.
If you are using a cable that has external metal armor or braid, terminate the external
armor or braid so that it does not pass through the gland grommet to ensure that the gland
can seal properly around the smooth, round inner cable jacket.
4.
Insert the cable through the PG16 gland.
The PG16 gland comes with a 5 mm to 9 mm grommet.
5.
Open the AC power input spring terminals by inserting a 2.5 mm flat-blade screwdriver
and pressing the lever.
6.
Connect the line, neutral, and protective earth ground connections by inserting wires into
the spring terminals on the AC power input terminals. Refer to the pinout below.
The AC power input terminals work with wiring up to 12 AWG. Use wiring that meets
your applicable electrical codes. You do not need to use ferrules.