Tools needed:
•
19 mm deep socket or wrench
•
22 mm socket or wrench
•
Ethernet termination tool
•
(Optional) 19 mm gland dome nut tightening tool (SEALCON part number S-1900-WR)
Complete the following steps to connect an Ethernet cable to the MON-10496.
Note
The RJ45 connector will not fit through the gland, so you must first pass the
Ethernet cable without the connector attached through the gland then terminate the
connector to the cable.
1.
Thread the M16 gland into the device enclosure.
2.
Tighten the gland body to a torque of 3.16 N · m (28 lb · in.).
3.
Install the backing nut onto the M16 gland and tighten to a torque of 2.5 N · m
(22.1 lb · in.).
4.
If you are using a cable that has external metal armor or braid, terminate the external
armor or braid so that it does not pass through the gland grommet to ensure that the gland
can seal properly around the smooth, round inner cable jacket.
5.
Insert the Ethernet cable through the gland using the appropriately sized grommet for
your cable.
The M16 gland comes with a pre-installed gray 2 mm to 6 mm grommet and an
additional black 4 mm to 8 mm grommet. For optimum sealing and strain relief, only use
wiring with a round, circular, and smooth exterior cable jacket.
6.
Terminate the RJ45 connector to the Ethernet cable and plug it into the Ethernet port.
7.
Tighten the M16 dome nut to a torque of 2.5 N · m (22.1 lb · in.) to provide strain relief
and weather resistance for the cable.
Connecting Power and Ground
Caution
Before opening the MON-10496 for service, you must turn off the
external power to the device and use the external breaker box to completely remove
mains power.
Attention
Avant d'ouvrir le MON-10496 pour l'entretien, vous devez couper
l'alimentation externe de l'appareil et utiliser le boîtier de disjoncteur externe pour
supprimer complètement l'alimentation secteur.
Hardware needed:
•
PG16 gland (supplied by NI) or 0.5 in. conduit fitting
•
5 mm to 9 mm diameter power cable with a smooth, round jacket and individual
conductors up to 12 AWG that comply with your local electrical code
12
|
ni.com
|
MON-10496 User Guide